DME1600
27
2
Messa
ge Ref
e
re
nce
Math Function Commands
There are eight math functions, but only one function can be enabled at a time. Each math
function either stores data for later use or performs mathematical operations on the mea-
sured values. These eight math functions are available to all measurement functions except
for continuity and diode testing.
The math functions use one or more internal registers. You can set the values in some of the
registers. The results of the math operations are stored in the registers.
CALC:FUNC
Sets the math function. Only one math function can be enabled at a time. The default setting
is for the percentage function to be enabled.
Command
CALCulate:FUNCtion {PERCent|AVERage|NULL|LIMit|MXB|DB|DBM}
CALCulate:FUNCtion?
Response
In response to CALC:FUNC?, the DME1600 returns PERC, AVER, NULL, LIM, MXB, DB, or DBM.
CALC:STAT
Turns the selected math function on or off. The default setting is “OFF.”
Command
CALCulate:STATe {OFF|ON}
CALCulate:STATe?
Response
In response to CALC:STAT?, the DME1600 returns “0” (OFF) or “1” (ON).
CALC:PERC:TARG
Sets the target value for the percentage math function. This value is cleared when you turn
the maximum and minimum math function off, turn off the power, or reset the remote interface
settings.
Command
CALCulate:PERCent:TARGet {<value>|MINimum|MAXimum}
CALCulate:PERCent:TARGet? [MINimum|MAXimum]
Response
In response to CALC:PERC:TARG?, the DME1600 returns the target value for the
percentage math function.
CALC:AVER:MIN
Returns the minimum value detected by the maximum and minimum math function. This
value is cleared when you turn the maximum and minimum math function off, turn off the
power, or reset the remote interface settings.
Command
CALCulate:AVERage:MINimum?
CALC:AVER:MAX
Returns the maximum value detected by the maximum and minimum math function. This
value is cleared when you turn the maximum and minimum math function off, turn off the
power, or reset the remote interface settings.
Command
CALCulate:AVERage:MAXimum?
Содержание DME1600
Страница 4: ...4 DME1600 This page left blank intentionally...
Страница 5: ...Setup This chapter explains the settings that you need to configure to use the interfaces 1...
Страница 9: ...Message Reference This chapter explains the SCPI com mands 2 Message Reference...
Страница 43: ...Appendix A Error Messages B Sample Programs...
Страница 55: ......