![Kikusui BIM1000 Series Скачать руководство пользователя страница 82](http://html.mh-extra.com/html/kikusui/bim1000-series/bim1000-series_user-manual_1966754082.webp)
82
User’s Manual
BIM1000 Series
Specifications | Basic specifications
Resistance meter
Sampling time
Judgment function
Item
Specifications
Measurement method
Four-terminal measurement method
Range
3 mΩ, 30 mΩ, 300 mΩ, 3 Ω, AUTO
Maximum display
value
*1
*1.
Displays OVER when the measurement range is exceeded.
3 mΩ range
3.1000 mΩ
30 mΩ range
31.000 mΩ
300 mΩ range
310.00 mΩ
3 Ω range
3.1000 Ω
Resolution
3 mΩ range
0.1 μΩ
30 mΩ range
1 μΩ
300 mΩ range
10 μΩ
3 Ω range
100 μΩ
Measured current
*2
*2.
Within error ± 10 %.
3 mΩ range
100 mA
30 mΩ range
100 mA
300 mΩ range
10 mA
3 Ω range
1 mA
Measurement frequency
1 kHz ±0.2 Hz
Accuracy
*3
*3.
Add ±3 digit when the sampling speed is set to FAST and ±2 digit when the sampling speed is set to MID.
±(0.5 % of r 5 digit)
Temperature coeffi-
cient
3 mΩ range
±(0.05 % of r 1 digit) /°C
30 mΩ range
±(0.05 % of r 0.5 digit) /°C
300 mΩ range
±(0.05 % of r 0.5 digit) /°C
3 Ω range
±(0.05 % of r 0.5 digit) /°C
Response time
*4
*4.
The time for the product’s measurement circuit to stabilize when a probe in an open state is connected to the DUT.
Approx. 2 ms
Item
Specifications
Sampling speed
Power supply frequency 50 Hz
FAST: 20 ms, MID: 50 ms, SLOW: 160 ms
Power supply frequency 60 Hz
FAST: 20 ms, MID: 42 ms, SLOW: 150 ms
Item
BIM1030
BIM1100
Judgment method
Window comparator method. Judgment made with software.
Resistance
Setting range
0.0000 Ω to 3.1000 Ω
Resolution
100 µΩ
Voltage
Setting range
0.000 V to 315.000 V
0.00 V to 1050.00 V
Resolution
1 mV
10 mV
Содержание BIM1000 Series
Страница 93: ......