BIM1000 Series
User’s Manual
63
Remote Control | Message Reference
Voltage measurement range
When you send a voltage value you want to measure, the range is set appropriately for measuring that
voltage.
Auto measurement range
Zero adjustment
Executes a zero adjustment and queries the result. Zero adjustment takes time. About 10 seconds are
required to the next process.
Clears zero adjustment
Program header
:VOLTage:RANGe
:VOLTage:RANGe?
Value
BIM1030
<0 to 315 (NR2)>
BIM1100
<-1050 to 1050 (NR2)>
Unit
V
Response
<NR3>
Example
Send “: VOLT: RANG 500”
The range is set to 600 V, which allows 500 V to be measured.
Send “VOLT: RANG?”
The present range setting “6.02” (600 V) is returned.
Note
If :AUTorange is set to ON (1) or VOLT, the value changes automatically.
Program header
:AUTorange
:AUTorange?
Value
{1|0|ON|OFF|RESistance|VOLTage}
{1|ON}:
Sets the resistance and voltage measurement range to auto
{0|OFF}:
Sets the resistance and voltage measurement range to manual
RESistance:
Sets the resistance measurement range to auto
VOLTage:
Sets the voltage measurement range to auto
Response
{ON|OFF|RESISTANCE|VOLTAGE}
Program header
:ADJust?
Response
{0|1}
0: Zero adjustment complete
1: Zero adjustment failure
Program header
:ADJust:CLEAr