BIM1000 Series
User’s Manual
81
Specifications
Unless specified otherwise, the specifications are for the following settings and conditions.
• The product is warmed up for at least 30 minutes.
• TYP: These are typical values that are representative of situations where the product operates in an environment
with an ambient temperature of 23 °C (73.4 °F). These values do not guarantee the performance of this product.
• setting: Indicates a setting.
• range: Indicates the rated value of each range.
• reading: Indicates a readout value.
Basic specifications
Voltmeter
Item
BIM1030
BIM1100
Rated input
±300 V
±1000 V
Range
6 V, 60 V, 300 V, AUTO
6 V, 60 V, 600 V, 1000 V, AUTO
Maximum display
value
*1
*1.
Displays OVER when the measurement range is exceeded.
6 V range
±6.30000 V
±6.30000 V
60 V range
±63.0000 V
±63.0000 V
300 V range
±315.000 V
–
600 V range
–
±630.000 V
1000 V range
–
±1050.00 V
Resolution
6 V range
10 μV
10 μV
60 V range
100 μV
100 μV
300 V range
1 mV
–
600 V range
–
1 mV
1000 V range
–
10 mV
Accuracy
*2
*2.
Add ±2 digit when the sampling speed is set to FAST or MID.
±(0.01 % of r 3 digit)
Temperature coefficient
±(0.001 % of r 0.3 digit) /°C
Response time
*3
*3.
The time for the product’s measurement circuit to stabilize when a probe in an open state is connected to the DUT.
Approx. 1 ms