Remote Control | Message Reference
*TRG
Measures once when :TRIGger:SOURce
is set to EXTernal. To execute TRG after changing the
measurement conditions, insert an interval of 100 ms.
Specific commands
Event status enable register
Sets or queries event status enable register 0 (ESER0) or event status enable register 1 (ESER1). Set
using a total value of the weight
of the bits you want to set.
Event status register
Queries event status register 0 (ESE0) or event status register 1 (ESE1). The set bits are returned using a
bit weight
.
QUEStionable status register
Queries events
Queries the event of the QUEStionable status register.
A query clears the contents of the register.
Program header
:ESE0
:ESE0?
:ESE1
:ESE1?
Value/Response
<0 to 255 (NR1)>
Example
• :ESE0 34
Set ESER0 bit 1 (bit weight 2) and bit 5 (bit weight 32) to 1.
• :ESE1? 34
ESER1 bit 1 (bit weight 2) and bit 5 (bit weight 32) are 1.
Program header
:ESR0?
:ESR1?
Response
<0 to 255 (NR1)>
Example
:ESR1? 34
ESER1 bit 1 (bit weight 2) and bit 5 (bit weight 32) are 1.
Program header
:STATus:QUEStionable[:EVENt]?
Response
<0 to 3845 (NR1)>
Example
:STATus:QUEStionable? 4
QUEStionable status register bit 2 (bit weight 4) is 1.