![Kikusui BIM1000 Series Скачать руководство пользователя страница 18](http://html.mh-extra.com/html/kikusui/bim1000-series/bim1000-series_user-manual_1966754018.webp)
18
User’s Manual
BIM1000 Series
Measurement | Measurement and Judgment Settings
Voltage measurement conditions
Set the measurement range, upper judgment limit, and lower judgment limit.
1
Press VOLTAGE.
2
Press the function key for the measurement condition you want to set, and select
the value.
The value switches each time you press the key.
Measurement condition
keys
Value
Description
Measurement range
(RANGE)
F1
AUTO, 6 V, 60 V,
300 V
*1
, 600 V
*2
,
1000 V
*1.
BIM1030 only
*2.
BIM1100 only
Select the measurement range.
Upper judgment limit
(UPPER)
F2
UPPER, OFF
If you specify UPPER, enter the upper judgment
limit using the numeric keypad, and then press
ENTER. When the measurement is greater than
or equal to the upper judgment limit, “UP”
appears in the display area.
Lower judgment limit
(LOWER)
F3
LOWER, OFF
If you specify LOWER, enter the lower judg-
ment limit using the numeric keypad, and then
press ENTER. When the measurement is less
than or equal to the lower judgment limit, “LOW”
appears in the display area.