BIM1000 Series
User’s Manual
33
External Control | Control and Monitor
When starting measurements with external trigger input
When the trigger source
is set to EXTERNAL, a measurement starts with a trigger input to TRIG-IN
(pin 20). The following figure shows a timing chart including the INDEX (pin 22) and EOM (pin 21) signal
outputs.
Monitoring judgment results
In resistance measurement, OHM-UP (pin 14) is at low level when the judgment result is UP and OHM-
LOW (pin 15) is at low level when the judgment result is LOW.
In voltage measurement, V-UP (pin 17) is at low level when the judgment result is UP and V-LOW (pin 18)
is at low level when the judgment result is LOW.
Item
Time
T1: Trigger input pulse width
1 ms or more
T2: Delay
The time set with delay
. 5 ms minimum.
T3: Measurement time
Measurement time × average count
Measurement time: Time set with the sampling speed
Average count: Average count set with the average function
T4: Computation time
0.3 ms
T5: EOM signal output (HOLD)
When EOM
is set to HOLD. Holds low level until the next TRIG-IN is
detected.
T6: EOM signal output (PULSE)
When EOM
is set to PULSE. Outputs a pulse for the specified time.
T7: Judgment delay
0.2 ms
Measurement
T1
ON
OFF
ON
ON
OFF
T2
T3
T4
T5
Delay
Judgment
result output
OFF
T6
ON
Delay
T7
TRIG-IN
INDEX
For HOLD
For PULSE
EOM
Computation