Connect a RF Flex cable between the Source IN and CPLR THRU port indicated in
, the
results should be similar to
.
Connect a RF Flex cable between the RCVR OUT and CPLR ARM port indicated in
the results should be similar to
Figure
61
Source IN to CPLR THRU & RCVR OUT to CPLR ARM Path Response
Table 24
Source Bypass Signal Path Test Instructions
Path #
Test Port Group
RF Path Description
Control Mode
<Address>.<Data>
Path
Components
1
Port 5, 9, 13 & 17
Source IN to CPLR THRU
0.0
S9
2
Port 6, 10, 14 & 18
Source IN to CPLR THRU
16.0
S11
3
Port 7, 11, 15 & 19
Source IN to CPLR THRU
32.0
S13
4
Port 8, 12, 16 & 20
Source IN to CPLR THRU
64.0
S15
Table 25
Receiver Bypass Signal Path Test Instructions
Path #
Test Port Group
RF Path Description
Control Mode
<Address>.<Data>
Path
Components
1
Port 5, 9, 13 & 17
RCVR OUT to CPLR ARM
0.0
S10
2
Port 6, 10, 14 & 18
RCVR OUT to CPLR ARM
16.0
S12
3
Port 7, 11, 15 & 19
RCVR OUT to CPLR ARM
32.0
S14
4
Port 8, 12, 16 & 20
RCVR OUT to CPLR ARM
64.0
S16