U3042AE16
Troubleshooting the Test Set
Source Signal Path Insertion Loss Test
Connect a RF Flex cable between the Test Port and Source IN port indicated in
,
the expected results should be similar to
Table 22
Source Signal Path Test Instructions
Path #
RF Path
Description
Control Mode
<Address>.<Data>
Path Components
Insertion Loss
(typical)
1
Source IN to Port 5
1
1. Use the Source IN port associated with this group of test ports.
0.1
P5 CPLR, S1 & 9
2
Source IN to Ports9
0.2
P9 CPLR, S1 & 9
3
Source IN to Port 13
0.3
P13 CPLR, S1 & 9
4
Source IN to Port 17
0.4
P17 CPLR, S1 & S9
5
Source IN to Port 6
16.1
P6 CPLR, S3& 11
6
Source IN to Port10
16.2
P10 CPLR, S11
7
Source IN to Port 14
16.3
P14 CPLR, S11
8
Source IN to Port 18
16.4
P18 CPLR, S3 & S11
9
Source IN to Port 7
32.1
P7 CPLR, S5 & 13
10
Source IN to Port 11
32.2
P11 CPLR, S11
11
Source IN to Port 15
32.3
P15 CPLR, S11
12
Source IN to Port 19
32.4
P19 CPLR, S9 & S13
13
Source IN to Port 8
64.1
P8 CPLR, S7 & 15
14
Source IN to Port 12
64.2
P12 CPLR, S11
15
Source IN to Port 16
64.3
P16 CPLR, S11
16
Source IN to Port 20
64.4
P20 CPLR, S7 & S15
72
User's and Service Guide U3042-90012