![Keysight U2741A Скачать руководство пользователя страница 90](http://html.mh-extra.com/html/keysight/u2741a/u2741a_programmers-reference-manual_1961514090.webp)
6
SENSe Subsystem
90
Keysight U2741A Programmer’s Reference Guide
[SENSe:]FRESistance:RANGe
Syntax
[SENSe:]FRESistance:RANGe <range|MIN|MAX|AUTO>
This command sets the measurement range for four-wire resistance
measurements.
[SENSe:]FRESistance:RANGe?
This command queries the measurement range for four-wire resistance
measurements.
Parameter
Remarks
– Selecting a fixed range will disable autorange.
– If the input signal is greater than can be measured on the selected range
(manual range), the U2741A gives an overload indication of
+9.9E+37
from
the remote interface.
– The U2741A is set to 1 k
Ω
, with autorange enabled, after a factory reset (
*RST
command).
Examples
FRES:RANG 100
//Configures the U2741A to measure
resistance at 100
Ω
range
FRES:RANG?
//Queries the resistance range
Typical Response:
+1.02
See Also
[SENSe:]FRESistance:RANGe:AUTO
Item
Type
Range of values
Default value
range
Numeric
100
Ω
, 1 k
Ω
, 10 k
Ω
, 100 k
Ω
, 1 M
Ω
, 10 M
Ω
, 100 M
Ω
MIN = 100
Ω
MAX = 100 M
Ω
1 k
Ω
Содержание U2741A
Страница 1: ...Keysight U2741A USB Modular 5 5 Digits Digital Multimeter Programmer s Reference Guide ...
Страница 7: ...Keysight U2741A Programmer s Reference Guide 7 Calibration Errors 164 13 Factory Default Settings ...
Страница 8: ...8 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 9: ...Keysight U2741A Programmer s Reference Guide 9 List of Figures Figure 2 1 Status system diagram 20 ...
Страница 10: ...10 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 12: ...12 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 112: ...6 SENSe Subsystem 112 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 126: ...7 STATus Subsystem 126 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 136: ...9 TRIGger Subsystem 136 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 154: ...11 IEEE Common Commands 154 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...