11
IEEE Common Commands
148
Keysight U2741A Programmer’s Reference Guide
*OPC/*OPC?
Syntax
*OPC
Sets the “Operation Complete” bit (bit 0) in the Standard Event register after all of
the previous commands have been completed.
*OPC?
This query always returns 1 to the output buffer after the previous commands
have been completed. Other commands cannot be executed until this command
completes.
Examples
*OPC
//This command sets the “Operation
Complete” bit (bit 0)
*OPC?
//The following query returns 1 after the
operation has completed
Typical Response:
1
Содержание U2741A
Страница 1: ...Keysight U2741A USB Modular 5 5 Digits Digital Multimeter Programmer s Reference Guide ...
Страница 7: ...Keysight U2741A Programmer s Reference Guide 7 Calibration Errors 164 13 Factory Default Settings ...
Страница 8: ...8 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 9: ...Keysight U2741A Programmer s Reference Guide 9 List of Figures Figure 2 1 Status system diagram 20 ...
Страница 10: ...10 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 12: ...12 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 112: ...6 SENSe Subsystem 112 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 126: ...7 STATus Subsystem 126 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 136: ...9 TRIGger Subsystem 136 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 154: ...11 IEEE Common Commands 154 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...