3
CALibrate Subsystem
28
Keysight U2741A Programmer’s Reference Guide
CALibration:ADC?
Syntax
CALibration:ADC?
This command performs a low-level calibration of the analog-to-digital (ADC)
converter circuitry. This is the first step in a recalibration sequence, and must be
done before doing the offset calibration.
Remarks
It will return a
+0
if the calibration passes and
+1
if it fails. Refer to the
Keysight
U2741A USB Modular 5.5 Digits Digital Multimeter Service Guide
for detailed
procedures, including how to connect to a calibration source, recommended
equipments, the specified calibration points, and so forth.
Example
CAL:ADC?
//Commands the U2741A to execute the
calibration
Typical Response:
+0
See Also
Содержание U2741A
Страница 1: ...Keysight U2741A USB Modular 5 5 Digits Digital Multimeter Programmer s Reference Guide ...
Страница 7: ...Keysight U2741A Programmer s Reference Guide 7 Calibration Errors 164 13 Factory Default Settings ...
Страница 8: ...8 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 9: ...Keysight U2741A Programmer s Reference Guide 9 List of Figures Figure 2 1 Status system diagram 20 ...
Страница 10: ...10 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 12: ...12 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 112: ...6 SENSe Subsystem 112 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 126: ...7 STATus Subsystem 126 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 136: ...9 TRIGger Subsystem 136 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 154: ...11 IEEE Common Commands 154 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...