![Keysight U2741A Скачать руководство пользователя страница 140](http://html.mh-extra.com/html/keysight/u2741a/u2741a_programmers-reference-manual_1961514140.webp)
10
Other Commands
140
Keysight U2741A Programmer’s Reference Guide
INITiate[:IMMediate]
Syntax
INITiate[:IMMediate]
This command changes the state of the triggering system from the “idle” state to
the “wait-for-trigger” state. Measurements will begin when the specified trigger
conditions are satisfied following the receipt of the
INIT
command.
Remarks
– To retrieve the reading, use the
FETC?
command.
– The
ABOR
command may be used to set the U2741A to idle state.
– The
INIT
command only initiates the measurement and needs a trigger (
BUS
or
IMMediate
) to make the actual measurement.
Examples
CONF 10, 0.00003
//Configures to measure DC voltage at
10 V range and 30
μ
V resolution
TRIG:SOUR IMM
//Sets the trigger source to immediate
INIT
//Places the U2741A in
wait-for-trigger state
FETC?
//The FETCh? command transfers the
reading from memory to the U2741A
output buffer
Typical Response:
+1.00
See Also
NOTE
The INIT command also clears the previous reading from the memory.
Содержание U2741A
Страница 1: ...Keysight U2741A USB Modular 5 5 Digits Digital Multimeter Programmer s Reference Guide ...
Страница 7: ...Keysight U2741A Programmer s Reference Guide 7 Calibration Errors 164 13 Factory Default Settings ...
Страница 8: ...8 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 9: ...Keysight U2741A Programmer s Reference Guide 9 List of Figures Figure 2 1 Status system diagram 20 ...
Страница 10: ...10 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 12: ...12 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 112: ...6 SENSe Subsystem 112 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 126: ...7 STATus Subsystem 126 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 136: ...9 TRIGger Subsystem 136 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 154: ...11 IEEE Common Commands 154 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...