SCPI Command Terminators
A command string sent to the U2741A must terminate with a <
new line
> (
<NL>
)
character. The IEEE-488 End-Of-Identify (EOI) message is interpreted as a
<NL>
character and can be used to terminate a command string in place of a
<NL>
character. A <carriage return> followed by a
<NL>
is also accepted. Command
string termination will always reset the current SCPI command path to the root
level.
IEEE-488.2 Common Commands
The IEEE- 488.2 standard defines a set of common commands that perform
functions such as reset, self-test, and status operation. Common commands
always begin with an asterisk ( * ), are three characters in length, and may include
one or more parameters. The command keyword is separated from the first
parameter by a blank space. Use a semicolon ( ; ) to separate multiple commands
as shown below.
*RST; *CLS; *ESE 32; *OPC?
Using the MIN, MAX, and AUTO Parameters
You can substitute
MINimum, MAXimum
, or
AUTOmatic
in place of a parameter for
many commands. For example, consider the following command:
SENS:VOLT:DC:RANG [<range|MINimum|MAXimum|AUTOmatic>]
Instead of selecting a specific voltage range, you can substitute
MIN
to set the range to its minimum value,
MAX
to set the range to its
maximum value, or
AUTO
to set it to automatic.
Содержание U2741A
Страница 1: ...Keysight U2741A USB Modular 5 5 Digits Digital Multimeter Programmer s Reference Guide ...
Страница 7: ...Keysight U2741A Programmer s Reference Guide 7 Calibration Errors 164 13 Factory Default Settings ...
Страница 8: ...8 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 9: ...Keysight U2741A Programmer s Reference Guide 9 List of Figures Figure 2 1 Status system diagram 20 ...
Страница 10: ...10 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 12: ...12 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 112: ...6 SENSe Subsystem 112 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 126: ...7 STATus Subsystem 126 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 136: ...9 TRIGger Subsystem 136 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 154: ...11 IEEE Common Commands 154 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...