![Keysight U2741A Скачать руководство пользователя страница 134](http://html.mh-extra.com/html/keysight/u2741a/u2741a_programmers-reference-manual_1961514134.webp)
9
TRIGger Subsystem
134
Keysight U2741A Programmer’s Reference Guide
TRIGger:SOURce
Syntax
TRIGger:SOURce <source>
This command selects the trigger source for measurements. The U2741A will
accept a software (bus) command, immediate (continuous) trigger, or a star
trigger.
TRIGger:SOURce?
This command queries the trigger source setting.
Parameter
Remarks
– After selecting the trigger source, you must place the U2741A in the
“wait-for-trigger” state using the
INIT
command. A trigger will not be
accepted from the selected trigger source until the U2741A is in the
“wait-for-trigger” state. The
INIT
command only initiates the measurement
and needs a trigger (
BUS
or
IMMediate
) to make the actual measurement.
– For the
IMMediate
(continuous) source, the trigger signal is always present.
When you place the U2741A in the “wait-for-trigger” state, the trigger is
issued immediately.
– For the
BUS
(software) source, the U2741A is triggered by the
*TRG
command
received over the remote interface. The
*TRG
command will not be accepted
unless the multimeter is in the “wait-for-trigger” state.
– A
CONF
or
MEAS?
command overwrites the
BUS
trigger and sets the trigger
source to
IMMediate
.
– The U2741A selects the immediate trigger source after a factory reset (
*RST
command).
– The bus trigger mode is initiated by sending a bus trigger command (
*TRG
),
after selecting
BUS
as the trigger source.
– The star trigger is only applicable when the U2741A is used in the U2781A
chassis. It is use for triggering multiple units in the chassis.
Item
Type
Range of values
Default value
source
Discrete
{IMMediate|BUS|STRG}
IMM
Содержание U2741A
Страница 1: ...Keysight U2741A USB Modular 5 5 Digits Digital Multimeter Programmer s Reference Guide ...
Страница 7: ...Keysight U2741A Programmer s Reference Guide 7 Calibration Errors 164 13 Factory Default Settings ...
Страница 8: ...8 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 9: ...Keysight U2741A Programmer s Reference Guide 9 List of Figures Figure 2 1 Status system diagram 20 ...
Страница 10: ...10 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 12: ...12 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 112: ...6 SENSe Subsystem 112 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 126: ...7 STATus Subsystem 126 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 136: ...9 TRIGger Subsystem 136 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 154: ...11 IEEE Common Commands 154 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...