2
SCPI Status Registers
22
Keysight U2741A Programmer’s Reference Guide
Standard Event Register
The Standard Event register reports the following types of instrument events:
power-on detected, command syntax errors, command execution errors,
calibration errors, query errors, or when an
*OPC
command is executed. Any or all
of these conditions can be reported in the Standard Event summary bit through
enable register. To set the enable register mask, you write a decimal value to the
register using the event status enable (
*ESE
) command.
Bit Definitions: Standard Event Register
The Standard Event register is cleared when:
– You execute the clear status (
*CLS
) command.
– Querying the event register using the event status register (
*ESR?
) command.
Bit number
Decimal value
Definition
0 Operation Complete
1
All commands prior to and including
*OPC
have been executed
1 Not Used
2
Always zero
2 Query Error
4
The U2741A tried to read the output buffer but it was empty. Or, a
new command line was received before a previous query has been
read. Or, both the input and output buffers are full.
3 Device Error
8
A self-test or calibration error occurred. For a complete listing of
the error messages, see
.
4 Execution Error
16
An execution error occurred (an error in the –200 range has been
generated)
5 Command Error
32
A command syntax error occurred (an error in the –100 range has
been generated)
6 Not Used
64
Always zero
7 Power On
128
Power has been turned off and on since the last time the event
register was read or cleared
Содержание U2741A
Страница 1: ...Keysight U2741A USB Modular 5 5 Digits Digital Multimeter Programmer s Reference Guide ...
Страница 7: ...Keysight U2741A Programmer s Reference Guide 7 Calibration Errors 164 13 Factory Default Settings ...
Страница 8: ...8 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 9: ...Keysight U2741A Programmer s Reference Guide 9 List of Figures Figure 2 1 Status system diagram 20 ...
Страница 10: ...10 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 12: ...12 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 112: ...6 SENSe Subsystem 112 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 126: ...7 STATus Subsystem 126 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 136: ...9 TRIGger Subsystem 136 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 154: ...11 IEEE Common Commands 154 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...