Step 22. Remove Option 028 License

Procedure Requirements

— The analyzer must be powered up and operating to perform this procedure.
— The Network Analyzer program must be running.
— A keyboard and mouse must be connected to the network analyzer.

Option 028 License Removal Procedure

1.

To start the Keysight License Manager, press Start > Keysight License 

Manager > Keysight License Manager . An Keysight License Manager 

dialog box will appear.

2.

Right click the on the desired option and click 

Delete

.

3.

In the Select Desired Option list, click 028. 

4.

In the Keysight License Manager dialog box that appears, press or click 

Yes

 to confirm delete.

5.

A message displays stating that the option removal was successful.

6.

Restart the PNA Analyzer application: Press 

File

 > 

Exit

.

7.

In the Exit NA Application dialog box that opens, press 

OK

.

Step 23. Enable Option 029

Procedure Requirements

— The analyzer must be powered up and operating to perform this procedure.
— The Network Analyzer program must be running.
— A keyboard and mouse must be connected to the network analyzer.
— Refer to the license message you received from Keysight: Verify that the 

analyzer’s model and serial numbers match those on the license message 

you received from Keysight.

— Obtain your license key file(s) for installation of this upgrade by following 

the instructions on the supplied Software Entitlement Certificate.

If Option 28 is not loaded on your PNA, proceed to 

“Step 23. Enable Option 

029” on page 44

For this step, you will need a USB flash drive.

Содержание PNA-X N5241B

Страница 1: ... PNA X N5241B Option 217 Option 219 Option 222 or Option 224 N5242B Option 217 Option 219 Option 222 or Option 224 N5249A Option 217 Option 219 Option 222 or Option 224 to include Option 029 Upgrade Kit Order Number N5241BU 229 N5242BU 229 and N5249BU 229 Keysight Kit Number N5242 60120 ...

Страница 2: ...U S government customers under its standard commercial license which is embodied in its End User License Agreement EULA a copy of which can be found at http www keysight com find sweula The license set forth in the EULA represents the exclusive authority by which the U S government may use modify distribute or disclose the Software The EULA and the license set forth therein does not require or per...

Страница 3: ...Keysight on page 5 IMPORTANT This kit contains the newer N5245 60124 noise receiver board with tabs If your instrument serial prefix is less than 5200 you need to order the N5245 60125 noise receiver board kit Contact Keysight to swap out the N5245 60124 board for a N5245 60125 kit Refer to Contacting Keysight on page 5 operating system The following may apply to your B mode PNA In June 2013 the N...

Страница 4: ...ure measurement capability to your Option 217 Option 219 Option 222 or Option 224 2 port analyzer by adding Option 029 which includes a noise down converter and noise receiver a splitter to provide a signal to the noise down converter receiver a bypass switch to each source port channel ...

Страница 5: ...e Web at http www keysight com find assist If you do not have access to the Internet please contact your Keysight field engineer If You Have Problems With the Upgrade Kit Contents Keysight stands behind the quality of the upgrade kit contents If you have problems with any item in the kit refer to www keysight com and the Contact Keysight link In any correspondence or telephone conversation refer t...

Страница 6: ...cense key s file from the Keysight Software Manager http www keysight com find softwaremanager To complete the request you will need to gather the following information From the certificate Order number Certificate number From your instrument The PNA contains extremely sensitive components that can be ruined if mishandled Follow instructions carefully when making cable connections especially wire ...

Страница 7: ...com 2 In the Search box enter the model number of the analyzer Ex N5242A and click Search 3 Click Technical Support Manuals 4 Click Service Manual 5 Click the service guide title to download the PDF file 6 When the PDF of the Service Guide is displayed scroll through the Contents section bookmarks to locate the information needed Protecting Your Workspace from Electrostatic Discharge For informati...

Страница 8: ...in lbs 1 N A 1 in torque wrench set to 72 in lbs 1 N A Use a 5 16 in torque wrench set to 10 in lbs on all cable connections except the front and rear panel bulkhead connectors Torque these connections to 21 in lb Products affected N5242B N5242B and N5249B Options 217 219 222 or 224 Installation to be performed by Keysight service center or personnel qualified by Keysight Estimated installation ti...

Страница 9: ... A54 Bridge 1 5087 7794 A55 Noise downconverter 1 5087 7767 Hardware Miscellaneous Switch attenuator bracket 2 N5242 00031 Bias tee bracket 2 N5242 00032 Machine screw M3 0 x 18 pan head to attach A52 and A53 to switch brackets 4 0515 0666 Machine screw M3 0 x 6 pan head to attach bias tee brackets 4 0515 0430 Machine screw M3 0 x 20 pan head to attach A54 to side frame 1 0515 1410 Hex nut with lo...

Страница 10: ...to A42 port 1 receiver attenuator 1 N5242 20277 W133 RF cable A42 port 1 receiver attenuator to A23 mixer brick A 1 N5242 20275 W134 RF cable A45 port 2 receiver attenuator to A23 mixer brick B 1 N5242 20276 W135 RF cable front panel Port 2 RCVR B IN to A45 port 2 receiver attenuator 1 N5242 20278 W136 RF cable A33 reference mixer switch to A23 mixer brick R1 1 N5242 20274 W137 RF cable A28 port 2...

Страница 11: ...fore proceeding with this installation Extra quantities of items such as protective plastic caps screws cable ties and cable clamps may be included in this upgrade kit It is normal for some of these items to remain unused after the upgrade is completed This installation requires the removal of the analyzer s protective outer covers The analyzer must be powered down and disconnected from the mains ...

Страница 12: ...d Install the A54 Bridge and A55 Noise Downconverter Step 10 Install the A22 Splitter Instruments with Serial Number Prefixes 5200 Step 10 Install the A22 Splitter All Instruments Step 12 Install the A7 Noise Receiver Board Step 13 Install the Bulkhead Connectors in the Test Set Front Plate For All Analyzers with Serial Numbers Prefixed MY SG US5310 and Below Step 14 Install the Test Set Cables St...

Страница 13: ...l number serial number or option number do not match those on your license key you will not be able to install the option If this is the case contact Keysight for assistance before beginning the installation of this upgrade Refer to Contacting Keysight on page 5 Once the license key file has been received and the information verified you can proceed with the installation at step 2 If the model num...

Страница 14: ...zer 3 Remove the foot locks item from the four bottom feet item and then remove the four bottom feet from the outer cover 4 Remove the four rear panel feet item by removing the center screws item with a T 20 TORX driver 5 Slide the outer cover toward the rear of the analyzer and remove it Figure 1 Outer Cover Removal This procedure is best performed with the analyzer resting on its front handles i...

Страница 15: ...s Step 3 Remove the Inner Cover Refer to Figure 2 for this step of procedure 1 With a T 10 TORX driver remove the 12 pan head screws item 2 With a T 10 TORX driver remove the 9 flat head screws item 3 Lift off the cover Figure 2 Inner Cover Removal ...

Страница 16: ...mpers item 2 With a T 10 TORX driver remove the screws item from the sides of the frame 3 Slide the front panel over the test port connectors 4 Disconnect the front panel interface ribbon cable item The front panel is now free from the analyzer Figure 3 Front panel Assembly Removal Before removing the front panel from the analyzer lift and support the front of the analyzer chassis ...

Страница 17: ...ort 2 bridge to front panel REF 2 SOURCE OUT W83 A28 port 2 bridge to A37 port 2 source attenuator W72 A34 port 1 source attenuator to front panel Port 1 SOURCE OUT W87 Front panel Port 1 RCVR A IN to A42 port 2 receiver attenuator W93 Front panel Port 2 RCVR B IN to A45 port 2 receiver attenuator W40 W154 in Option 224 PNA Front panel REF 2 RCVR R2 IN to A23 mixer brick R2 W84 A37 port 2 source a...

Страница 18: ...elow IMPORTANT save these cables for reuse W115 Front panel Port 1 CPLR ARM to A29 Port 1 coupler W116 A32 port 2 coupler to front panel Port 2 CPLR THRU W117 Front panel Port 2 CPLR ARM to A32 Port 2 coupler 3 Disconnect the ribbon cable from each of the step attenuator connectors It is not necessary to disconnect these ribbon cables from the A19 test set motherboard connector designations are gi...

Страница 19: ...above Remove two screws item from the A38 port 1 bias tee and the A41 port 2 bias tee and lift them out of the PNA and set them aside for re installation later For analyzers with serial numbers prefixed MY SG US5310 and below Remove two screws item from the A38 port 1 bias tee and the A41 port 2 bias tee and lift them out of the PNA These old bias tees have 2 4 mm connectors and will be replaced l...

Страница 20: ...20 Installation Note N5242 90120 All Instruments Figure 4 Existing Attenuators and Brackets Removal ...

Страница 21: ...re that the switches are oriented as shown they are not oriented the same for both ports The A52 switch with the longest cable is for port 1 2 Secure each switch to its bracket using two screws item for each 3 Remove the source and receiver attenuators from the old bracket and using the same screws item attach them to the new brackets Make sure that they are oriented as shown they are not oriented...

Страница 22: ...22 Installation Note N5242 90120 All Instruments Figure 5 Attenuators and Bypass Switches Assembly ...

Страница 23: ...th ports 2 Secure the attenuator brackets to the analyzer test set deck using three screws item for each If bias tee option is installed NOT N5241 42 9B 217 or 222 For analyzers with serial numbers prefixed MY SG US5321 and above Position the existing bias tees configured as shown in Figure 6 on the new bias tee brackets and secure each of them with the original two mounting screws item For analyz...

Страница 24: ...24 Installation Note N5242 90120 All Instruments Figure 6 New Brackets Installation into the Analyzer Step 9 Assemble and Install the A54 Bridge and A55 Noise Downconverter ...

Страница 25: ...n cable item to the A55 noise downconverter connector as shown 3 Place the combined A54 W131 A55 assembly into the analyzer as shown 4 Secure the A55 noise downconverter to the side frame using three screws item as shown 5 Secure the A54 bridge to the side frame using one screw item and one hex nut with lock washer item as shown Install the screw from the outside of the side frame through the top ...

Страница 26: ...is step of the procedure Although only Option 219 is shown in the illustration Option 217 222 and 224 are similar in appearance New parts are listed in Table 1 on page 9 1 Position the A22 splitter on top of the mixer brick mounting block as shown 2 Secure the A22 splitter with two screws item as indicated ...

Страница 27: ...Installation Note N5242 90120 27 All Instruments Figure 8 A22 Splitter Installation ...

Страница 28: ... the instructions in your N5245 60125 kit to modify your N5245 60124 to a no tabs noise board Refer to http literature cdn keysight com litweb pdf N5245 90125 pdf This section assumes you have an instrument with a serial prefix 5200 and have contacted Keysight to swap your N5245 60124 board for a N5245 60125 kit Refer to Contacting Keysight on page 5 If your instrument serial is 5200 skip to next ...

Страница 29: ...f this step of the procedure Although only Option 219 is shown in the illustration Option 217 222 and 224 are similar in appearance New parts are listed in Table 1 on page 9 1 Place the analyzer top side up on a flat surface 2 Insert the A7 noise receiver board in the analyzer as shown Make sure it is fully seated in the motherboard connector ...

Страница 30: ...30 Installation Note N5242 90120 All Instruments Figure 9 Noise Receiver Board Installation ...

Страница 31: ...e listed in Table 1 on page 9 1 From the back side of the test set front plate insert a bulkhead connector into a hole in the plate 2 Install 1x washer and 1x nut Hand tighten nut and ensure bulkhead connector hexagon nut on the back side of test set front plate is aligned to the test set subpanel hexagon indent 3 Repeat previous two steps for the remaining bulkhead connectors 4 Torque nuts on the...

Страница 32: ...he order listed Use a 5 16 in torque wrench set to 10 in lbs to tighten all cable connectors For all analyzer serial numbers W135 N5242 20278 Front panel Port 2 RCVR B IN to A45 port 2 receiver attenuator W129 N5242 20293 A53 port 2 bypass switch to A54 port 2 bridge W128 N5242 20303 A53 port 2 bypass switch to front panel Port 2 SOURCE OUT W130 N5242 20302 A53 port 2 bypass switch to A54 port 2 b...

Страница 33: ...242 20273 A25 port 1 bridge to A34 port 1 source attenuator W139 N5242 20301 A22 splitter to A55 noise down converter W134 N5242 20276 A45 port 2 receiver attenuator to A23 mixer brick B W138 N5242 20280 Front panel REF 2 RCVR R2 IN to A23 mixer brick R2 W43 N5222 20007 A22 splitter to A23 mixer brick W42 N5222 20009 A21 HMA26 5 to A22 splitter For analyzers with the bias tee option installed NOT ...

Страница 34: ...the test set deck to the top side of the analyzer The other ends will be connected in the next step W143 N5242 20300 Semi rigid cable A55 noise down converter to A7 noise receiver board RF W141 N5242 20299 Semi rigid cable A55 noise downconverter to A7 noise receiver board LO W142 N5242 60041 Flexible cable A55 noise downconverter J4 to A7 noise receiver board P2 Refer to Figure 11 for this part o...

Страница 35: ...Installation Note N5242 90120 35 All Instruments Figure 11 New Test Set Cable Installation Part 2 ...

Страница 36: ...s installed you need not replace it a The part number should be as shown in Table 1 on page 9 b If the part number IS as shown proceed to Step 16 Remove the Old Lower Front Panel Overlay c If the part number IS NOT as show continue with this step 2 Remove the three flat head screws item from the power supply bracket 3 Remove the seven pan head screws item from the power supply rear panel 4 Slide t...

Страница 37: ...Installation Note N5242 90120 37 All Instruments Figure 12 Power Supply Replacement ...

Страница 38: ...t in the cutouts in the lower front dress panel to push on the old overlay item and separate it from the front dress panel 2 From the front side of the front panel pull off the overlay completely and discard it 3 Remove any adhesive remaining on the front panel IMPORTANT To avoid possible damage to the lower front panel overlay do not attempt to attach the lower front panel label until Step 18 Ins...

Страница 39: ...Installation Note N5242 90120 39 All Instruments Figure 13 Lower Front panel Overlay Replacement ...

Страница 40: ...anel cable connectors have been tightened using a 5 16 in torque wrench set to 21 in lbs 2 Reconnect the ribbon cable item to the A1 front panel interface board 3 Slide the front panel over the front panel connectors 4 With a T 10 TORX driver reinstall the 12 screws item in the sides of the frame Figure 14 Front Panel Assembly Re installation Before installing the front panel assembly onto the ana...

Страница 41: ...inst the edges of the recess 3 Once the overlay is in place press it firmly onto the frame to secure it s Step 19 Position the Cables and Wires to Prevent Pinching On the top side of the PNA carefully position the gray flex cables so they can t be pinched between the covers and the rails On the bottom side of the PNA carefully fold or push down the ribbon cables and wires so they can t be pinched ...

Страница 42: ...he Inner Cover Refer to Figure 15 for this step of the procedure 1 Position the inner cover on the analyzer 2 With a T 10 TORX driver install the 12 pan head screws item 3 With a T 10 TORX driver install the 9 flat head screws item Figure 15 Inner Cover re installation ...

Страница 43: ...ver 3 Install the four bottom feet item onto the bottom of the outer cover then install the foot locks item 4 Install the strap handles item by tightening the screws item on both ends of each strap handle with a T 20 TORX driver Figure 16 Outer Cover re installation This procedure is best performed with the analyzer resting on its front handles in the vertical position Do not place the analyzer on...

Страница 44: ...m delete 5 A message displays stating that the option removal was successful 6 Restart the PNA Analyzer application Press File Exit 7 In the Exit NA Application dialog box that opens press OK Step 23 Enable Option 029 Procedure Requirements The analyzer must be powered up and operating to perform this procedure The Network Analyzer program must be running A keyboard and mouse must be connected to ...

Страница 45: ... be installed The instrument will not prompt you when the upgrade has been completed 6 The PNA application automatically restarts after the option installation has been completed Optional Cycle power on the instrument to activate the newly installed option upgrade s Press the power button once to power down the instrument Option Verification Procedure Once the analyzer has restarted and the Networ...

Страница 46: ... Chapter 3 bookmark Tests and Adjustments in the PDF Service Guide1 After the specified adjustments have been performed the analyzer should operate and phase lock over its entire frequency range Operator s Check Perform the Operator s Check to check the basic functionality of the analyzer For instructions click the Chapter 3 bookmark Tests and Adjustments in the PDF Service Guide1 If you experienc...

Страница 47: ... 1 If necessary reinstall front jumper cables 2 If necessary reinstall the cable guards pushing them over the front jumper cables until the cushioning material touches the front panel of the PNA 3 Install the dust caps on the test ports 4 Clean the analyzer as needed using a damp cloth ...

Страница 48: ...48 Installation Note N5242 90120 All Instruments ...

Страница 49: ...Installation Note Xxxxx xxxxx 3 ...

Страница 50: ...This information is subject to change without notice Keysight Technologies 2012 2017 Edition 1 May 2017 N5242 90120 www keysight com ...

Отзывы: