Installation Note N5242-90120 43
All Instruments
Step 21. Reinstall the Outer Cover
Refer to
for this step of the procedure.
1. Slide the outer cover over the analyzer frame.
2. Install the four rear panel feet (item
⑤
) by installing the center screws
(item
⑥
) with a T-20 TORX driver,.
3. Install the four bottom feet (item
④
) onto the bottom of the outer cover
then install the foot locks (item
③
).
4.
Install the strap handles (item
①
) by tightening the screws (item
➁
) on
both ends of each strap handle with a T-20 TORX driver.
Figure 16
Outer Cover re-installation
This procedure is best performed with the analyzer resting on its front
handles in the vertical position.
Do not place the analyzer on its front panel
without the handles
. This will damage the front panel assemblies.
Содержание PNA-X N5241B
Страница 20: ...20 Installation Note N5242 90120 All Instruments Figure 4 Existing Attenuators and Brackets Removal ...
Страница 22: ...22 Installation Note N5242 90120 All Instruments Figure 5 Attenuators and Bypass Switches Assembly ...
Страница 27: ...Installation Note N5242 90120 27 All Instruments Figure 8 A22 Splitter Installation ...
Страница 30: ...30 Installation Note N5242 90120 All Instruments Figure 9 Noise Receiver Board Installation ...
Страница 35: ...Installation Note N5242 90120 35 All Instruments Figure 11 New Test Set Cable Installation Part 2 ...
Страница 37: ...Installation Note N5242 90120 37 All Instruments Figure 12 Power Supply Replacement ...
Страница 39: ...Installation Note N5242 90120 39 All Instruments Figure 13 Lower Front panel Overlay Replacement ...
Страница 48: ...48 Installation Note N5242 90120 All Instruments ...
Страница 49: ...Installation Note Xxxxx xxxxx 3 ...