Installation Note N5242-90120 33
All Instruments
- W127(N5242-20292) A37 port 2 source attenuator to A53 port 2 bypass
switch
- W132(N5242-20277) Front-panel Port 1 RCVR A IN to A42 port 1
receiver attenuator
- W124(N5242-20295) Front-panel Port 1 CPLR THRU to A52 port 1
bypass switch
- W122(N5242-20298) A34 port 1 source attenuator to A52 port 1 bypass
switch
- W123(N5242-20297) A52 port 1 bypass switch to front-panel Port 1
SOURCE OUT
- W126(N5242-20272) A28 port 2 bridge to A37 port 2 source attenuator
- W137(N5242-20279) A28 port 2 bridge to front-panel REF 2 SOURCE
OUT
- W136(N5242-20274) A33 reference mixer switch to A23 mixer brick (R1)
- W133(N5242-20275) A42 port 1 receiver attenuator to A23 mixer brick
(A)
- W121(N5242-20273) A25 port 1 bridge to A34 port 1 source attenuator
- W139(N5242-20301) A22 splitter to A55 noise down converter
- W134(N5242-20276) A45 port 2 receiver attenuator to A23 mixer brick
(B)
- W138(N5242-20280) Front-panel REF 2 RCVR R2 IN to A23 mixer brick
(R2)
- W43 (N5222-20007) A22 splitter to A23 mixer brick
- W42 (N5222-20009) A21 HMA26.5 to A22 splitter
For analyzers with the bias tee option installed (NOT N5241/42/9B-217
or N5241/42/9B-222):
All analyzer serial numbers:
- W125 (N5242-20296) A52 port 1 bypass switch to A38 port 1 bias tee
- W85 N5222-20035) front-panel Port 2 CPLR THRU to A41 port 2 bias
tee
- W119 (N5222-20072) A38 port 1 bias tee to A29 port 1 coupler
- W120 (N5222-20065) A41 port 2 bias tee to A32 port 2 coupler
Analyzer serial numbers prefixed MY/SG/US5310 and below:
- W117(N5222-20071) A32 port 2 coupler to front-panel Port 2 CPLR
ARM
- W115(N5222-20070) Front-panel Port 1 CPLR ARM to A29 Port 1
coupler
Содержание PNA-X N5241B
Страница 20: ...20 Installation Note N5242 90120 All Instruments Figure 4 Existing Attenuators and Brackets Removal ...
Страница 22: ...22 Installation Note N5242 90120 All Instruments Figure 5 Attenuators and Bypass Switches Assembly ...
Страница 27: ...Installation Note N5242 90120 27 All Instruments Figure 8 A22 Splitter Installation ...
Страница 30: ...30 Installation Note N5242 90120 All Instruments Figure 9 Noise Receiver Board Installation ...
Страница 35: ...Installation Note N5242 90120 35 All Instruments Figure 11 New Test Set Cable Installation Part 2 ...
Страница 37: ...Installation Note N5242 90120 37 All Instruments Figure 12 Power Supply Replacement ...
Страница 39: ...Installation Note N5242 90120 39 All Instruments Figure 13 Lower Front panel Overlay Replacement ...
Страница 48: ...48 Installation Note N5242 90120 All Instruments ...
Страница 49: ...Installation Note Xxxxx xxxxx 3 ...