34 Installation Note N5242-90120
All Instruments
-
For analyzers with the bias tee option not installed (N5241/42/9B-217 or
N5241/42/9B-222):
All analyzer serial numbers:
- W145 (N5242-20305) A52 port 1 bypass switch to A29 port 1 coupler
Analyzer serial numbers prefixed MY/SG/US5310 and below:
- W115 (N5222-20070) Front-panel Port 1 CPLR ARM to A29 Port 1
coupler
- W116 (N5222-20069) Front-panel Port 2 CPLR THRU to A32 Port 2
coupler
- W117 (N5222-20071) Front-panel Port 2 CPLR ARM to A32 Port 2
coupler
3. Position the analyzer as shown in
(fans facing
upwards) and loosely install the following cables. Route each of the cables
through the opening in the test set deck to the top side of the analyzer.
The other ends will be connected in the next step.
- W143(N5242-20300) Semi-rigid cable, A55 noise down converter to A7
noise receiver board RF
- W141(N5242-20299) Semi-rigid cable, A55 noise downconverter to A7
noise receiver board LO
- W142(N5242-60041) Flexible cable, A55 noise downconverter J4 to A7
noise receiver board P2
Refer to
for this part of this step of the procedure. Although only
Option 219 is shown in the illustration, Option 224 is similar in appearance.
New parts are listed in
4. The analyzer should be positioned on its left side (fans facing upwards) as
shown.
5. Connect semi-rigid cables W141 and W143 as indicated. Torque
connectors to 10 in-lbs.
6. Connect flexible cable W142 as indicated.
7. Go back and torque the connectors on the other ends of W141 and W143
to 10 in-lbs.
Содержание PNA-X N5241B
Страница 20: ...20 Installation Note N5242 90120 All Instruments Figure 4 Existing Attenuators and Brackets Removal ...
Страница 22: ...22 Installation Note N5242 90120 All Instruments Figure 5 Attenuators and Bypass Switches Assembly ...
Страница 27: ...Installation Note N5242 90120 27 All Instruments Figure 8 A22 Splitter Installation ...
Страница 30: ...30 Installation Note N5242 90120 All Instruments Figure 9 Noise Receiver Board Installation ...
Страница 35: ...Installation Note N5242 90120 35 All Instruments Figure 11 New Test Set Cable Installation Part 2 ...
Страница 37: ...Installation Note N5242 90120 37 All Instruments Figure 12 Power Supply Replacement ...
Страница 39: ...Installation Note N5242 90120 39 All Instruments Figure 13 Lower Front panel Overlay Replacement ...
Страница 48: ...48 Installation Note N5242 90120 All Instruments ...
Страница 49: ...Installation Note Xxxxx xxxxx 3 ...