44 Installation Note N5242-90120
All Instruments
-
Step 22. Remove Option 028 License
Procedure Requirements
— The analyzer must be powered up and operating to perform this procedure.
— The Network Analyzer program must be running.
— A keyboard and mouse must be connected to the network analyzer.
Option 028 License Removal Procedure
1.
To start the Keysight License Manager, press Start > Keysight License
Manager > Keysight License Manager . An Keysight License Manager
dialog box will appear.
2.
Right click the on the desired option and click
Delete
.
3.
In the Select Desired Option list, click 028.
4.
In the Keysight License Manager dialog box that appears, press or click
Yes
to confirm delete.
5.
A message displays stating that the option removal was successful.
6.
Restart the PNA Analyzer application: Press
File
>
Exit
.
7.
In the Exit NA Application dialog box that opens, press
OK
.
Step 23. Enable Option 029
Procedure Requirements
— The analyzer must be powered up and operating to perform this procedure.
— The Network Analyzer program must be running.
— A keyboard and mouse must be connected to the network analyzer.
— Refer to the license message you received from Keysight: Verify that the
analyzer’s model and serial numbers match those on the license message
you received from Keysight.
— Obtain your license key file(s) for installation of this upgrade by following
the instructions on the supplied Software Entitlement Certificate.
If Option 28 is not loaded on your PNA, proceed to
.
For this step, you will need a USB flash drive.
Содержание PNA-X N5241B
Страница 20: ...20 Installation Note N5242 90120 All Instruments Figure 4 Existing Attenuators and Brackets Removal ...
Страница 22: ...22 Installation Note N5242 90120 All Instruments Figure 5 Attenuators and Bypass Switches Assembly ...
Страница 27: ...Installation Note N5242 90120 27 All Instruments Figure 8 A22 Splitter Installation ...
Страница 30: ...30 Installation Note N5242 90120 All Instruments Figure 9 Noise Receiver Board Installation ...
Страница 35: ...Installation Note N5242 90120 35 All Instruments Figure 11 New Test Set Cable Installation Part 2 ...
Страница 37: ...Installation Note N5242 90120 37 All Instruments Figure 12 Power Supply Replacement ...
Страница 39: ...Installation Note N5242 90120 39 All Instruments Figure 13 Lower Front panel Overlay Replacement ...
Страница 48: ...48 Installation Note N5242 90120 All Instruments ...
Страница 49: ...Installation Note Xxxxx xxxxx 3 ...