LED Indicators
The M9602A/M9603A module has two LED indicators, Access and Status. These
indicators show the status of the module. See
for the meanings. If the
emergency shutdown occurs, both indicators turn red.
Table 1-3
LED Status
Measurement Terminals
The M9602A/M9603A has the following measurement terminals. You can make
2-wire connection using the High force and Low force terminals or 4-wire
connection (Kelvin connection) using the High force, Low force, High sense and
Low sense terminals. The Kelvin connection is effective in high current
measurement.
LED
LED state
Status
Access
Solid Green
The module is turned on, the initialization is
completed, and the operation is ready.
Flashing Green
The “Identify PXIe module” feature is on.
Solid Red
Emergency shutdown occurs.
Off
The control session is closed.
Status
Solid Green
Output is enabled and the output voltage is within
42 V.
Solid Yellow
Output is enabled and the output voltage is over
42 V.
Solid Red
Emergency shutdown occurs.
Off
Output is disabled.