Keysight M9602A/M9603A Startup Guide, Edition 3
29
Connecting a DUT
Floating
When the M9602A/M9603A is shipped from the factory, the Low force terminal is
connected to the frame/chassis terminal by the short bar. The short bar must be
connected to perform the grounded measurement.
If you want to perform the floating measurement or if you want to connect another
instrument’s ground to the Low force, remove the short bar, and leave the
terminals open. Then use the Low force terminal on the HF or HS connector to
connect the other instrument.
For the floating measurement, do not apply voltage over
40 V to the Low force
terminal.
V
floating
, the voltage between Low force and the frame/chassis ground,
is limited to
40 V. Failure to heed this caution may result in damage to the
M9602A/M9603A.
Do not apply current or voltage to the frame/chassis terminal. Doing so will
damage the M9602A/M9603A.
To prevent electrical shock, do not touch any of measurement circuit at any time
while a floating measurement is in progress. Also use accessories that comply
with IEC 61010-031. All terminals and the extended conductors must be isolated
by using insulation caps, sleeves, etc.
Afin d’éviter toute décharge électrique, ne touchez aucune mesure de circuit à
tout moment lorsque la mesure de flotte est en cours. Utiliser également des
accessoires qui sont conformes à la norme IEC 61010-031. Toutes les bornes et
les conducteurs prolongés doivent être isolés en utilisant des bouchons
d'isolation, des manchons, etc.