Keysight M9602A/M9603A Startup Guide, Edition 3

Function Overview

Aperture time

 is the time required to acquire the measurement data. For accurate 

and reliable measurement, the aperture should be increased. This value can be 

specified.

Overhead time

 includes other factors such as range change or data 

compensation. This value depends on the measurement conditions and cannot be 

specified. 

Remote Transient Voltage Measurement

The M9602A/M9603A has a dedicated voltmeter with higher bandwidth for the 

remote transient voltage measurement. This function is effective in measuring 

pulse voltage with the width of a few to dozens of microseconds. You can enable 

or disable this function.

Automatic Output On/Off

The M9602A/M9603A module has output on/off settings when the trigger system 

changes the status. You can enable or disable these settings independently.

Auto Output On 

The source output is automatically turned on just before the 

trigger system is initiated by programming.
When the trigger system is initiated on the Soft Front Panel, 

this setting is ignored.

Auto Output Off 

The source output is automatically turned off when all trigger 

systems change the status from busy to idle.

Output Off Mode

The output off mode is the condition immediately after the source output is turned 

off. The mode should be specified by programming before enabling the source 

output. The following settings are available for the output off mode.

High impedance 

Output switch: off
Voltage source setup is not changed.
The current source setup is not changed if the current output 

value and range are less than or equal to 10 mA and 10 mA 

range. Otherwise, the output value is set to 10 mA, and the 

output range is set to 10 mA range.

Normal 

Output switch: off
Source function: voltage source

Содержание M9602A PXIe

Страница 1: ...Keysight M9602A M9603A PXIe Precision Source Measure Unit Startup Guide...

Страница 2: ...oftware under the same terms by which the software is customarily provided to the public Accordingly Keysight provides the Software to U S government customers under its stan dard commercial license w...

Страница 3: ...trument in such an environment constitutes a definite safety hazard DO NOT REMOVE COVERS No operator serviceable parts inside Refer servicing to qualified personnel To prevent electrical shock do not...

Страница 4: ...AUTION notice until the indicated conditions are fully understood and met WARNING A WARNING denotes a hazard It calls atten tion to an operating procedure or practice that if not correctly performed o...

Страница 5: ...rn the Module for Service 10 Step 2 Verify Shipment Contents 11 Step 3 Install the Software 12 Requirements 12 Installation Process 12 Step 4 Install the Module 13 Step 5 Verify Operation of the Modul...

Страница 6: ...35 Calibration 36 Function Overview 37 Limit Compliance 37 Operation Mode 37 Output Filter 38 Pulse Output 38 Sweep Output 39 List Sweep 39 Ranging Mode 39 Seamless Current Measurement Ranging 40 Trig...

Страница 7: ...also supplies software drivers that allow you to support the modules in many popular PXI chassis and programming environments You can control the M9602A M9603A using IVI drivers LabVIEW driver is also...

Страница 8: ...age or destroy electronic components Use a static safe work station to perform all work on electronic assemblies This figure shows a static safe work station using two types of ESD protection Conducti...

Страница 9: ...t you can install the module soon after unpacking Refer to the PXIe chassis manual for chassis installation and details Inspect for Damage After unpacking a module carefully inspect it for any shippin...

Страница 10: ...or example M9603A Product serial number for example MYXXXXXXXX The serial number label is located on the side of the module Description of failure or service required 4 Carefully pack the module in it...

Страница 11: ...A M9603A are shown below Table 1 2 Consumable Supplies Model number Description Quantity M9602A or M9603A PXIe Source Measure Unit 15 MSa s 60 V 3 5 A DC 10 5 A pulse 1 NA Quick Startup Poster 1 NA Ce...

Страница 12: ...refer to the PXIe chassis documentation Minimum 1 5 GB free space is required on the controller for installing the M9602A M9603A software Installation Process Visit www keysight com find m9602a or ww...

Страница 13: ...o from the chassis power down the chassis to prevent damage to the module Place the chassis on a static safe work station see Electrostatic Discharge ESD Precautions on page 8 and then position it so...

Страница 14: ...s into the front module guides top and bottom Slide the module to the rear of the chassis and assure that the injector ejector handle is pushed down in the unlatched downward position Slide the module...

Страница 15: ...f you remove the module from the chassis Immediately after powering down the M9602A M9603A module side panels may be hot to cause a burn Before removing the module from the chassis take a time for dis...

Страница 16: ...m the following procedure to check the communication between the module and the controller 1 Make sure that the chassis and the controller are turned off 2 If you are using a remote controller connect...

Страница 17: ...t Panel the M960x SFP 3 Confirm that the Access indicator on the M9602A M9603A front panel turns green Refer to Figure 1 1 and Table 1 3 for LED status 4 Select Firmware Update from the Utilities menu...

Страница 18: ...M9603A modules to connect and click Connect to launch the M960x SFP 3 Confirm that the Access indicator on the M9602A M9603A front panel turns green Refer to Figure 1 1 and Table 1 3 for LED status 4...

Страница 19: ...e M9602A M9603A front panel LED Indicators Measurement Terminals Trigger and Interlock Terminals Figure 1 1 M9602A M9603A Front Panel Access indicator Status indicator Low sense High force High sense...

Страница 20: ...and Low force terminals or 4 wire connection Kelvin connection using the High force Low force High sense and Low sense terminals The Kelvin connection is effective in high current measurement LED LED...

Страница 21: ...erminal when the shielding box access door is opened If you change the connection interface test fixture prober and such connect an interlock cable to the one actually used Confirm periodically that t...

Страница 22: ...e la porte d acc s la protection de la bo te est ouverte Si vous changez l interface de connexion un test d installation la sonde ou toute autre l ment connectez un cordon d enclenchement celui utilis...

Страница 23: ...ge which appears in High force High sense and Guard is expressed by the equation 60 6 V Vfloating Where Vfloating is the voltage between Low force and the frame chassis ground Vfloating is 0 V in the...

Страница 24: ...t or if you want to connect another instrument s ground to the Low force remove the short bar and leave the terminals open Then use the Low force terminal on the HF or HS connector to connect the othe...

Страница 25: ...The input trigger should be the TTL level Intlk Interlock terminals These terminals are used to connect an interlock circuit which is installed in your test fixture or shielding box The interlock cir...

Страница 26: ...charge and heat to prevent electrical shock and burn Use gloves and any tool Also have enough time for discharge and radiation Afin d viter de toucher l extr mit du c ble d allongement ou l aire de l...

Страница 27: ...have a negative impact on the performance of the instrument Use coaxial cables and shielding technique to minimize the impact 2 Wire Connection or 4 Wire Connection When connecting a device under tes...

Страница 28: ...gure 1 3 2 Wire Connection and 4 Wire Connection A V High force 2 wire connection Low force Frame chassis A V High force 4 wire connection Low force Frame chassis High sense Low sense Short bar Short...

Страница 29: ...ltage between Low force and the frame chassis ground is limited to 40 V Failure to heed this caution may result in damage to the M9602A M9603A Do not apply current or voltage to the frame chassis term...

Страница 30: ...core and the inner shield Therefore ideally the current measured by the instrument is the same as the current at the DUT terminal because there is no leakage current For several connection examples s...

Страница 31: ...interlock circuit is important and necessary to prevent electrical shock when the user touches the measurement terminals The limit value or Interlock threshold voltage level is programmable within 42...

Страница 32: ...t 2 ea Connector terminal block 2 5 mm 6 terminal 1 ea BNC to ferrule terminal cable Keysight PX0101A 001 1 5 m PX0101A 002 3 0 m or equivalent 1 ea BNC connector jack to soldering terminals 1 ea Conn...

Страница 33: ...ct the mechanical switches directly to the connector terminal block by connection wires and ferrule terminals Figure 1 6 Dimension of the Interlock Switch Keysight N1254A 402 3 1 35 6 8 1 14 NC NO COM...

Страница 34: ...etergents or chemical solvents Do not attempt to clean the inside To clean the connectors use alcohol in a well ventilated area Allow all residual alcohol moisture to evaporate and the fumes to dissip...

Страница 35: ...t M9602A M9603A provides the self calibration function to maintain the measurement performance If the environmental temperature changes 5 C or more perform the self calibration This is effective in th...

Страница 36: ...indicator on the M9602A M9603A front panel turns green Refer to Figure 1 1 and Table 1 3 for LED status 4 Select Self Calibration from the Utilities menu on the soft front panel to open the Self calib...

Страница 37: ...ss Current Measurement Ranging Trigger System Measurement Time Remote Transient Voltage Measurement Automatic Output On Off Output Off Mode Interlock Protection from Emergency Limit Compliance Limit o...

Страница 38: ...er The M9602A M9603A module has a built in output filter to obtain clean output signal without spikes or overshoots However using a filter may increase the settling time You can enable or disable this...

Страница 39: ...output step The source output and the measurement are performed at the specified interval The minimum interval is 4 s Ranging Mode Output Range Source Range The following ranging modes are available f...

Страница 40: ...nt Measurement Ranging This function allows you to measure current with the best resolution over multiple ranges without changing the range In this function current measurement is performed at multipl...

Страница 41: ...nterval The timing is controlled by the software Depending on the controller the interval may be larger than the specified value TIMER Triggers at the specified interval The timing is controlled by th...

Страница 42: ...hanges the status You can enable or disable these settings independently Auto Output On The source output is automatically turned on just before the trigger system is initiated by programming When the...

Страница 43: ...oltage level when the interlock terminals are open The interlock threshold voltage level is programmable within 42 V and the default value is 42 V Set the threshold voltage to the voltage you think st...

Страница 44: ...ch as power supply failure or internal communication failure is detected the emergency shutdown occurs The module is powered off and two LED indicators turn red When the emergency including the over v...

Страница 45: ......

Страница 46: ...This information is subject to change without notice Keysight Technologies 2020 2021 Edition 3 May 2021 M9603 90000 M9603 90000 www keysight com...

Отзывы: