Keysight M9602A/M9603A Startup Guide, Edition 3
Step 5: Verify Operation of the Module
To Perform Self Test
You can perform the self test using the M960x PXIe Source Measure Unit Soft
Front Panel (the M960x SFP).
Before performing the self test, turn the instrument output off and disconnect
cables from the measurement terminals.
1. Click
Start menu > Keysight M960x Source Measure Unit > M960x SFP
. The
Connect to Instrument dialog box will open.
2. In the list on the window, highlight the M9602A/M9603A modules to connect,
and click
Connect
to launch the M960x SFP.
3. Confirm that the Access indicator on the M9602A/M9603A front panel turns
green. Refer to
and
for LED status.
4. Select
Self Test...
from the Utilities menu. The Self Test dialog box will open.
5. Click
Run Self Test
at the bottom of the Self Test dialog box to start the self
test. When the self test is completed, the results will be displayed in the Self
Test dialog box.
If the module fails the self test, it will be defective. See
to return the modules to Keysight.
Use the module within the rating. If you input a signal above the rating, the
module will be damaged.