Keysight M9602A/M9603A Startup Guide, Edition 3
21
Front Panel Features
There are potentially hazardous voltages (
60.6 V) present at the High force,
High sense, and Guard terminals of this instrument. To prevent electrical shock,
the following safety precautions must be observed during the use of the
instrument.
•
Use a three-conductor AC power cord to connect the cabinet (if used) and
the instrument to an electrical ground (safety ground).
•
If an interlock circuit is not installed in your test fixture or connection
interface, you must install and connect the interlock circuit that opens the
interlock terminal when the shielding box access door is opened.
•
If you change the connection interface, test fixture, prober, and such, connect
an interlock cable to the one actually used.
•
Confirm periodically that the interlock function works normally.
•
Before touching the connections on the High force, High sense, and Guard
terminals, turn the instrument off and discharge any capacitors. If you do
not
turn the instrument off, complete
all
of the following items, regardless of the
instrument settings.
•
Disable the output, and confirm that the Status indicator turns off.
•
Confirm that the Status indicator does not turn yellow.
•
Open the fixture cover or the shielding box access door (open interlock).
•
Discharge any capacitors connected to a channel.
•
Warn persons working around the instrument about dangerous conditions.