Safety Information

The following general safety precau-

tions must be observed during all 

phases of operation of this instrument. 

Failure to comply with these precau-

tions or with specific warnings or oper-

ating instructions in the product 

manuals violates safety standards of 

design, manufacture, and intended use 

of the instrument. Keysight Technolo-

gies assumes no liability for the cus-

tomer's failure to comply with these 

requirements. 

General

Do not use this product in any manner not 

specified by the manufacturer. The protec-

tive features of this product must not be 

impaired if it is used in a manner specified in 

the operation instructions.

Before Applying Power

Verify that all safety precautions are taken. 

Make all connections to the unit before 

applying power. Note the external markings 

described under “Safety Symbols”.

Ground the Instrument

Keysight chassis’ are provided with a 

grounding-type power plug. The 

instrument chassis and cover must be 

connected to an electrical ground to 

minimize shock hazard. The ground pin 

must be firmly connected to an electri-

cal ground (safety ground) terminal at 

the power outlet. Any interruption of 

the protective (grounding) conductor 

or disconnection of the protective 

earth terminal will cause a potential 

shock hazard that could result in per-

sonal injury.

Do Not Operate in an Explosive 

Atmosphere

Do not operate the module/chassis in 

the presence of flammable gases or 

fumes.

Do Not Operate Near Flammable 

Liquids

Do not operate the module/chassis in 

the presence of flammable liquids or 

near containers of such liquids.

Cleaning

Clean the outside of the Keysight mod-

ule/chassis with a soft, lint-free, 

slightly dampened cloth. Do not use 

detergent or chemical solvents.

Do Not Remove Instrument Cover

Only qualified, service-trained person-

nel who are aware of the hazards 

involved should remove instrument 

covers. Always disconnect the power 

cable and any external circuits before 

removing the instrument cover.

Keep away from live circuits

Operating personnel must not remove 

equipment covers or shields. Proce-

dures involving the removal of covers 

and shields are for use by service-

trained personnel only. Under certain 

conditions, dangerous voltages may 

exist even with the equipment 

switched off. To avoid dangerous elec-

trical shock, DO NOT perform proce-

dures involving cover or shield removal 

unless you are qualified to do so.

DO NOT operate damaged 

equipment

Whenever it is possible that the safety 

protection features built into this prod-

uct have been impaired, either through 

physical damage, excessive moisture, 

or any other reason, REMOVE POWER 

and do not use the product until safe 

operation can be verified by service-

trained personnel. If necessary, return 

the product to an Keysight Technolo-

gies Sales and Service Office for ser-

vice and repair to ensure the safety 

features are maintained.

DO NOT block the primary 

disconnect

The primary disconnect device is the 

appliance connector/power cord when 

a chassis used by itself, but when 

installed into a rack or system the dis-

connect may be impaired and must be 

considered part of the installation.

Do Not Modify the Instrument

Do not install substitute parts or per-

form any unauthorized modification to 

the product. Return the product to an 

Keysight Sales and Service Office to 

ensure that safety features are main-

tained.

In Case of Damage

Instruments that appear damaged or 

defective should be made inoperative 

and secured against unintended oper-

ation until they can be repaired by 

qualified service personnel

Do NOT block vents and fan exhaust: 

To ensure adequate cooling and venti-

lation, leave a gap of at least 50mm 

(2") around vent holes on both sides of 

the chassis. 
Do NOT operate with empty slots: To 

ensure proper cooling and avoid dam-

aging equipment, fill each empty slot 

with an AXIe filler panel module.
Do NOT stack free-standing chassis: 

Stacked chassis should be rack-

mounted. 
All modules are grounded through the 

chassis: During installation, tighten 

each module's retaining screws to 

secure the module to the chassis and 

to make the ground connection.

Operator is responsible to maintain 

safe operating conditions. To ensure 

safe operating conditions, modules 

should not be operated beyond the full 

temperature range specified in the 

Environmental and physical specifica-

tion. Exceeding safe operating condi-

tions can result in shorter lifespan, 

improper module performance and 

user safety issues. When the modules 

are in use and operation within the 

specified full temperature range is not 

maintained, module surface tempera-

tures may exceed safe handling condi-

tions which can cause discomfort or 

burns if touched. In the event of a 

module exceeding the full temperature 

range, always allow the module to cool 

before touching or removing modules 

from the chassis.

 

iv

 

Содержание M9521A

Страница 1: ...Keysight M9521A AXIe System Module ASM Startup Guide...

Страница 2: ......

Страница 3: ...rights in excess of these rights customarily provided to the public to use modify reproduce release per form display or disclose commercial computer software or commercial com puter software document...

Страница 4: ...erform proce dures involving cover or shield removal unless you are qualified to do so DO NOT operate damaged equipment Whenever it is possible that the safety protection features built into this prod...

Страница 5: ...at this ISM device complies with the Canadian ICES 001 This symbol represents the time period during which no hazardous or toxic substance elements are expected to leak or deteriorate during normal us...

Страница 6: ...vi...

Страница 7: ...First then the Host Controller 19 Step 6 Install IO Libraries Suite and Connect to the Web Interface 21 Host PC Requirements 21 Installing the Keysight IO Libraries Suite 22 Using Keysight Connection...

Страница 8: ...viii Keysight M9521A AXIe System Module Startup Guide...

Страница 9: ...the ASM and other installed modules appropriate control software and instrument drivers must be installed on the host PC This Startup Guide plus additional information can be found in PDF format on th...

Страница 10: ...pression 4 IN Accepts an externally sourced timing input Input range 5V peak AC coupled 50 terminated 250 mV minimum swing 5 OUT Extends the internal clock source to external instruments 3 3V AC coupl...

Страница 11: ...icant level of ESD protection Of the two only the table mat and wrist strap combination provides adequate ESD protection when used alone To ensure user safety the static safe accessories must provide...

Страница 12: ...hat you ordered This printed document Keysight M9521A AXIe System Module Startup Guide An AXIe Chassis and Controller Software and Product Information CD M9514 10001 This CD contains software drivers...

Страница 13: ...hen fully open the handles by pivoting them out towards you until they are perpendicular to the front panel The left handle is shown below from the top view correctly extended 5 Slide the ASM complete...

Страница 14: ...ectricity To minimize electrostatic damage take the necessary anti static precautions Some AXIe chassis provide a grounding terminal on the front panel to which you can connect a wrist strap Empty Slo...

Страница 15: ...re details The following table indicates the various states colors of the ASM Status LED and its general meaning LED State color Description M9521A ASM Status LED off ASM not receiving power Chassis p...

Страница 16: ...ed state or circuit is working normally Refer to the M9514A Startup Guide or M9514A and M9521A User Guide for information on the chassis STATUS LEDs Verify the STATUS LEDs on your ASM and the AXIe cha...

Страница 17: ...nframe a current list can be found at www keysight com find axie chassis If you decide to use an embedded controller module as your host PC you need to install the controller in the chassis before pro...

Страница 18: ...c connection For additional host PC and chassis ASM interface information refer to Host PC Chassis Interface Topics on page 28 of this guide Also refer to the M9514A and M9521A User Guide Except where...

Страница 19: ...ashing green you can turn on your PCIe connected controller You can also view the chassis web interface or bring up the chassis SFP ASM power high ASM powered up by Shelf Manager Chassis ASM and Exter...

Страница 20: ...e chassis must be powered down before modules are added or removed While the power sequencing shown above does not apply to an embedded controller because an embedded controller and chassis are powere...

Страница 21: ...equired software Microsoft NET Keysight IO Libraries version 17 1 or later and chassis device drivers the host PC requires the following Keysight s M9536A AXIe Embedded Controller requires Windows Emb...

Страница 22: ...properly enumerate PCIe devices in the AXIe chassis at www keysight com find axie chassis Keysight s M9536A Embedded Controller must have BIOS AG16 or later If you are using a Keysight embedded contr...

Страница 23: ...on on your desktop double click the icon to launch the program b From a Command Prompt window type c Program Files Keysight IO Libraries Suite KeysightConnectionExpert exe to launch the program This s...

Страница 24: ...cted Note also that the ASM soft front panel SFP provides significant additional functionality to the Web Interface However using the Web Interface requires only a browser on the host PC whereas using...

Страница 25: ...is a brief description of what you can do on each page ASM Firmware Version Home Page View General Information about the ASM such as product identification firmware version and LAN parameters LAN Conf...

Страница 26: ...configure On the M9514A LAN Configuration Web Page and the M9521A LAN Configuration Web Page there is a button labeled Reset User Settable Memory This button resets all user settable non volatile mem...

Страница 27: ...Information CD in the host PC If you do not have the CD or a CD drive on your PC you can download the most recent version of the SFP and drivers at www keysight com find M9521A 2 Select the M9521A me...

Страница 28: ...may be of assistance in establishing a connection between the host PC and the ASM Unless otherwise noted these topics relate to interfacing to the ASM itself not to other modules in the chassis slots...

Страница 29: ...ill be displayed by Keysight Connection Expert under the PXIO label not under the LAN label 5 For a PCIe connection to the chassis a private IP address beginning with 169 254 will be assigned If a dir...

Страница 30: ...only its LAN interface the PCIe to LAN network adapter will not be visible to your PC and the aforementioned driver does not need to be installed 8 If the chassis is connected only to LAN the chassis...

Страница 31: ...Keysight M9521A AXIe System Module Startup Guide 31 Host PC Chassis Interface Topics...

Страница 32: ...return address Product model number for example M9521A Product serial number for example TWXXXXXXXX The serial number label is located on the side of the module A description of failure or service req...

Страница 33: ...ng user replaceable assemblies and accessories are available from Keysight Only Keysight approved accessories shall be used Other parts originally supplied with the AXIe System Module or chassis may b...

Страница 34: ...sis do not have a system slot Environmental Operating and storage conditions Operating Storage Temperature 0 C to 50 C 25 C to 60 C Altitude up to 3000 meters up to 4600 meters Humidity Type tested at...

Страница 35: ...010 1 Canada CSA C22 2 No 61010 1 USA UL std no 61010 1 EMC Complies with European EMC Directive IEC EN 61326 1 CISPR Pub 11 Group 1 Class A AS NZS CISPR 11 ICES NMB 001 This ISM device complies with...

Страница 36: ...36 Keysight M9521A AXIe System Module Startup Guide Safety Related Specifications...

Страница 37: ...37...

Страница 38: ...This information is subject to change without notice Keysight Technologies 2017 Edition 10 December 2017 Published in USA M9521 90001 www keysight com...

Отзывы: