Step 3: Install the AXIe System Module in the Chassis
8
Retract the handle ends by sliding them outward on their metal shafts, away
from each other, toward the chassis edge; this secures them out of the way of
test connections.
At this step, you should also install the AXIe embedded controller (such as the
Keysight M9536A) if you are using one. The embedded controller may be
installed in any chassis slot except slot 7; however, it is best to install it in outer
slots to minimize impact with any modules using the chassis local bus for
E-Keying. If you are using an external host PC, do not connect it to the ASM yet.
Proceed to Step 4 to power up the AXIe chassis. Verify that the chassis fans are
operating and free of obstructions that may restrict airflow.
Detailed module installation and removal instructions may be
found in the
M9514A and M9521A User Guide.
Static Electricity
—The components and connectors on modules
are sensitive to static electricity. To minimize electrostatic
damage, take the necessary anti-static precautions. Some AXIe
chassis provide a grounding terminal on the front panel, to which
you can connect a wrist strap.
Empty Slots
—Except for performing initial chassis verification or
troubleshooting, do not operate the chassis with empty slots.
Always insert a filler panel module or instrument module into
empty slots. This is especially important for the slots on either
side of an instrument module. This allows proper air flow and
cooling, and provides EMI shielding for the chassis and installed
components. Leaving slots empty can increase fan speed, raise
ambient noise, overheat components, and shut down modules.
ASM
—The AXIe System Module is integral to the operation of the
chassis. Except for troubleshooting purposes, do not remove the
ASM. The ASM is installed in slot 7 of the chassis.
Hot Swap
—AXIe does not explicitly support hot swap for
instrument modules. Keysight recommends fully powering down
the chassis before installing or removing modules
Embedded Controller Module
—Manually initiate a graceful
shutdown of the Windows controller operating system and power
off the chassis before removing the controller.
Содержание M9521A
Страница 1: ...Keysight M9521A AXIe System Module ASM Startup Guide...
Страница 2: ......
Страница 6: ...vi...
Страница 8: ...viii Keysight M9521A AXIe System Module Startup Guide...
Страница 31: ...Keysight M9521A AXIe System Module Startup Guide 31 Host PC Chassis Interface Topics...
Страница 36: ...36 Keysight M9521A AXIe System Module Startup Guide Safety Related Specifications...
Страница 37: ...37...