Keysight M9521A Скачать руководство пользователя страница 14

14

Keysight M9521A AXIe System Module Startup Guide

Step 3: Install the AXIe System Module in the Chassis

8

Retract the handle ends by sliding them outward on their metal shafts, away 

from each other, toward the chassis edge; this secures them out of the way of 

test connections.

At this step, you should also install the AXIe embedded controller (such as the 

Keysight M9536A) if you are using one. The embedded controller may be 

installed in any chassis slot except slot 7; however, it is best to install it in outer 

slots to minimize impact with any modules using the chassis local bus for 

E-Keying. If you are using an external host PC, do not connect it to the ASM yet.
Proceed to Step 4 to power up the AXIe chassis. Verify that the chassis fans are 

operating and free of obstructions that may restrict airflow. 

Detailed module installation and removal instructions may be 

found in the 

M9514A and M9521A User Guide.

Static Electricity

—The components and connectors on modules 

are sensitive to static electricity. To minimize electrostatic 

damage, take the necessary anti-static precautions. Some AXIe 

chassis provide a grounding terminal on the front panel, to which 

you can connect a wrist strap. 

Empty Slots

—Except for performing initial chassis verification or 

troubleshooting, do not operate the chassis with empty slots. 

Always insert a filler panel module or instrument module into 

empty slots. This is especially important for the slots on either 

side of an instrument module. This allows proper air flow and 

cooling, and provides EMI shielding for the chassis and installed 

components. Leaving slots empty can increase fan speed, raise 

ambient noise, overheat components, and shut down modules.

ASM

—The AXIe System Module is integral to the operation of the 

chassis. Except for troubleshooting purposes, do not remove the 

ASM. The ASM is installed in slot 7 of the chassis.

Hot Swap

—AXIe does not explicitly support hot swap for 

instrument modules. Keysight recommends fully powering down 

the chassis before installing or removing modules

Embedded Controller Module

—Manually initiate a graceful 

shutdown of the Windows controller operating system and power 

off the chassis before removing the controller.

Содержание M9521A

Страница 1: ...Keysight M9521A AXIe System Module ASM Startup Guide...

Страница 2: ......

Страница 3: ...rights in excess of these rights customarily provided to the public to use modify reproduce release per form display or disclose commercial computer software or commercial com puter software document...

Страница 4: ...erform proce dures involving cover or shield removal unless you are qualified to do so DO NOT operate damaged equipment Whenever it is possible that the safety protection features built into this prod...

Страница 5: ...at this ISM device complies with the Canadian ICES 001 This symbol represents the time period during which no hazardous or toxic substance elements are expected to leak or deteriorate during normal us...

Страница 6: ...vi...

Страница 7: ...First then the Host Controller 19 Step 6 Install IO Libraries Suite and Connect to the Web Interface 21 Host PC Requirements 21 Installing the Keysight IO Libraries Suite 22 Using Keysight Connection...

Страница 8: ...viii Keysight M9521A AXIe System Module Startup Guide...

Страница 9: ...the ASM and other installed modules appropriate control software and instrument drivers must be installed on the host PC This Startup Guide plus additional information can be found in PDF format on th...

Страница 10: ...pression 4 IN Accepts an externally sourced timing input Input range 5V peak AC coupled 50 terminated 250 mV minimum swing 5 OUT Extends the internal clock source to external instruments 3 3V AC coupl...

Страница 11: ...icant level of ESD protection Of the two only the table mat and wrist strap combination provides adequate ESD protection when used alone To ensure user safety the static safe accessories must provide...

Страница 12: ...hat you ordered This printed document Keysight M9521A AXIe System Module Startup Guide An AXIe Chassis and Controller Software and Product Information CD M9514 10001 This CD contains software drivers...

Страница 13: ...hen fully open the handles by pivoting them out towards you until they are perpendicular to the front panel The left handle is shown below from the top view correctly extended 5 Slide the ASM complete...

Страница 14: ...ectricity To minimize electrostatic damage take the necessary anti static precautions Some AXIe chassis provide a grounding terminal on the front panel to which you can connect a wrist strap Empty Slo...

Страница 15: ...re details The following table indicates the various states colors of the ASM Status LED and its general meaning LED State color Description M9521A ASM Status LED off ASM not receiving power Chassis p...

Страница 16: ...ed state or circuit is working normally Refer to the M9514A Startup Guide or M9514A and M9521A User Guide for information on the chassis STATUS LEDs Verify the STATUS LEDs on your ASM and the AXIe cha...

Страница 17: ...nframe a current list can be found at www keysight com find axie chassis If you decide to use an embedded controller module as your host PC you need to install the controller in the chassis before pro...

Страница 18: ...c connection For additional host PC and chassis ASM interface information refer to Host PC Chassis Interface Topics on page 28 of this guide Also refer to the M9514A and M9521A User Guide Except where...

Страница 19: ...ashing green you can turn on your PCIe connected controller You can also view the chassis web interface or bring up the chassis SFP ASM power high ASM powered up by Shelf Manager Chassis ASM and Exter...

Страница 20: ...e chassis must be powered down before modules are added or removed While the power sequencing shown above does not apply to an embedded controller because an embedded controller and chassis are powere...

Страница 21: ...equired software Microsoft NET Keysight IO Libraries version 17 1 or later and chassis device drivers the host PC requires the following Keysight s M9536A AXIe Embedded Controller requires Windows Emb...

Страница 22: ...properly enumerate PCIe devices in the AXIe chassis at www keysight com find axie chassis Keysight s M9536A Embedded Controller must have BIOS AG16 or later If you are using a Keysight embedded contr...

Страница 23: ...on on your desktop double click the icon to launch the program b From a Command Prompt window type c Program Files Keysight IO Libraries Suite KeysightConnectionExpert exe to launch the program This s...

Страница 24: ...cted Note also that the ASM soft front panel SFP provides significant additional functionality to the Web Interface However using the Web Interface requires only a browser on the host PC whereas using...

Страница 25: ...is a brief description of what you can do on each page ASM Firmware Version Home Page View General Information about the ASM such as product identification firmware version and LAN parameters LAN Conf...

Страница 26: ...configure On the M9514A LAN Configuration Web Page and the M9521A LAN Configuration Web Page there is a button labeled Reset User Settable Memory This button resets all user settable non volatile mem...

Страница 27: ...Information CD in the host PC If you do not have the CD or a CD drive on your PC you can download the most recent version of the SFP and drivers at www keysight com find M9521A 2 Select the M9521A me...

Страница 28: ...may be of assistance in establishing a connection between the host PC and the ASM Unless otherwise noted these topics relate to interfacing to the ASM itself not to other modules in the chassis slots...

Страница 29: ...ill be displayed by Keysight Connection Expert under the PXIO label not under the LAN label 5 For a PCIe connection to the chassis a private IP address beginning with 169 254 will be assigned If a dir...

Страница 30: ...only its LAN interface the PCIe to LAN network adapter will not be visible to your PC and the aforementioned driver does not need to be installed 8 If the chassis is connected only to LAN the chassis...

Страница 31: ...Keysight M9521A AXIe System Module Startup Guide 31 Host PC Chassis Interface Topics...

Страница 32: ...return address Product model number for example M9521A Product serial number for example TWXXXXXXXX The serial number label is located on the side of the module A description of failure or service req...

Страница 33: ...ng user replaceable assemblies and accessories are available from Keysight Only Keysight approved accessories shall be used Other parts originally supplied with the AXIe System Module or chassis may b...

Страница 34: ...sis do not have a system slot Environmental Operating and storage conditions Operating Storage Temperature 0 C to 50 C 25 C to 60 C Altitude up to 3000 meters up to 4600 meters Humidity Type tested at...

Страница 35: ...010 1 Canada CSA C22 2 No 61010 1 USA UL std no 61010 1 EMC Complies with European EMC Directive IEC EN 61326 1 CISPR Pub 11 Group 1 Class A AS NZS CISPR 11 ICES NMB 001 This ISM device complies with...

Страница 36: ...36 Keysight M9521A AXIe System Module Startup Guide Safety Related Specifications...

Страница 37: ...37...

Страница 38: ...This information is subject to change without notice Keysight Technologies 2017 Edition 10 December 2017 Published in USA M9521 90001 www keysight com...

Отзывы: