Performance Verification Tests
4
Keysight 3458A Calibration Manual
77
4
Subtract the 1 Hz Unit Under Test Readings from the 1 Hz Frequency Standard
Value. Record the difference on the
Test Card.
5
Change the Frequency Standard to 10 MHz and record the exact Frequency
Standard Value on the
Test Card. Execute Trig, and record
the Unit Under Test Reading on the
Test Card.
6
Subtract the 10 MHz Unit Under Reading from the 10 MHz Frequency
Standard Value. Record the difference on the
Test Card.
7
If either of the differences are beyond the limits specified, the instrument
should be adjusted. See
Chapter 3, “Adjustment Procedures”
adjustments.
Содержание 3458A
Страница 1: ...Keysight 3458A Multimeter Calibration Manual ...
Страница 10: ...10 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 14: ...14 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 16: ...16 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 18: ...18 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 39: ...Adjustment Procedures 3 Keysight 3458A Calibration Manual 39 Figure 3 1 4 terminal short ...
Страница 52: ...3 Adjustment Procedures 52 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 59: ...Performance Verification Tests 4 Keysight 3458A Calibration Manual 59 Figure 4 1 General test procedure ...
Страница 109: ...Command Summary 5 Keysight 3458A Calibration Manual 109 Related commands ACAL CAL SECURE ...
Страница 116: ...A Specifications 116 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...