Command Summary
5
Keysight 3458A Calibration Manual
113
TEMP?
Description
Temperature Query
. Returns the multimeter's internal temperature in degrees
Centigrade.
Syntax
TEMP?
Remarks
– Monitoring the multimeter's temperature is helpful to determine when to
perform autocalibration.
–
Related commands
: ACAL, CAL, CALSTR
Example
10 OUTPUT 722; "TEMP?"
!READ TEMPERATURE
20 ENTER 722; A
!ENTER RESULT
30 PRINT A
!PRINT RESULT
40 END