B
Electronic Calibration of the 3458A (Product Note 3458A-3)
122
Keysight 3458A Calibration Manual
Gain adjustments are now made for all other DC
voltage ranges.
4
Using the input path for the 10 V range,
accurately measure
1 V generated internally.
Linearity of the measurement circuits allows a
measurement that accurately reflects the actual l
V output. In other words, we transfer traceable
accuracy from the 10 V range to all other ranges.
The lower ranges use amplifiers to condition the
input for the 10 V full-scale analog-to-digital
converter. Each amplifier used requires a gain
constant, G
A
, to adjust normal readings. The
following process determines these constants.
5
In the 1 V range, measure the same 1 V
previously measured with the 10 V range.
6
Calculate a 1 V range gain adjustment so that
the two measurements agree. Note that
neither the precise value nor the long-term
stability of the internal 1 volt source are
important. The internal 1 volt source need
only be stable for the time it takes to measure
it twice.
7
Using the adjusted 1 V range, accurately
measure 0.1 V generated internally.
8
Measure the same 0.1 V using the 100 mV
range.
9
Calculate a 100 mV range gain adjustment so
that the two measurements agree.
Normal 100 V and 1000 V range measurements
use a 100:1 resistor network to attenuate the
input. To correct errors introduced by this
network, we apply zero volts to the input. Then,
we apply 10 V and measure the actual value.
Finally, we measure 0.1 V, with the zero error
removed, and compute the gain adjustment
constant.
Input voltages greater than 100 V (1000 V range)
create a self-heating error in the resistor network,
as shown in Figure 3. This easily identified error is
simply specified as part of the instrument's
published error. Additional measurements result
in constants to compensate for switching
transients and leakage currents.
Figure 3.
On the 1000 V range, nonlinear self-heating errors of the 100:1
resistive attenuator are noticeable.
Resistance and DC current
adjustments
Calibration of all resistance (nine ranges from
10
W
to 1 G
W
) and DC current ranges (eight
ranges from 100 nA to 1 A) requires only one
external resistance standard. Resistance is
measured by applying a known current through
the unknown resistance and measuring the
voltage across it. Current is measured by
applying the unknown current through a known
shunt resistor and measuring the voltage across
it. The process explained previously has already
corrected errors in the DC voltage input path.
Measuring the actual values of the current
sources and shunt resistors results in the
Содержание 3458A
Страница 1: ...Keysight 3458A Multimeter Calibration Manual ...
Страница 10: ...10 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 14: ...14 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 16: ...16 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 18: ...18 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 39: ...Adjustment Procedures 3 Keysight 3458A Calibration Manual 39 Figure 3 1 4 terminal short ...
Страница 52: ...3 Adjustment Procedures 52 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 59: ...Performance Verification Tests 4 Keysight 3458A Calibration Manual 59 Figure 4 1 General test procedure ...
Страница 109: ...Command Summary 5 Keysight 3458A Calibration Manual 109 Related commands ACAL CAL SECURE ...
Страница 116: ...A Specifications 116 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...