3
Adjustment Procedures
48
Keysight 3458A Calibration Manual
Figure 3-4
Characterizing the adjustment setup
4
Set the synthesizer to deliver a 3 V rms sinewave at a frequency of
2 MHz. Adjust the synthesizer's output voltage until the voltage displayed on
the multimeter is within 0.2% of the voltage recorded on Line A. Record the
synthesizer's voltage setting on Line C of the Adjustment Record.
5
Set the synthesizer to deliver a 3 V rms sinewave at a frequency of
8 MHz. Adjust the synthesizer until the voltage displayed on the multimeter is
within 0.2% of the voltage recorded on Line A. Record the synthesizer's
voltage setting on Line D of the Adjustment Record.
6
Set the synthesizer to deliver a 1 V rms sinewave at a frequency of
100 kHz. Replace the 3 V thermal converter with the 1 V thermal converter.
Record the exact DC voltage measured by the multimeter on Line E of the
Adjustment Record.
Содержание 3458A
Страница 1: ...Keysight 3458A Multimeter Calibration Manual ...
Страница 10: ...10 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 14: ...14 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 16: ...16 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 18: ...18 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 39: ...Adjustment Procedures 3 Keysight 3458A Calibration Manual 39 Figure 3 1 4 terminal short ...
Страница 52: ...3 Adjustment Procedures 52 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 59: ...Performance Verification Tests 4 Keysight 3458A Calibration Manual 59 Figure 4 1 General test procedure ...
Страница 109: ...Command Summary 5 Keysight 3458A Calibration Manual 109 Related commands ACAL CAL SECURE ...
Страница 116: ...A Specifications 116 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...