5
Command Summary
114
Keysight 3458A Calibration Manual
TEST
Description
Causes the multimeter to perform a series of internal self-test.
Syntax
TEST
Remarks
– Always disconnect any input signals before you run self-test. If you leave an
input signal connected to the multimeter, it may cause a self-test failure.
– If a hardware error is detected. the multimeter sets bit 0 in the error register
and a more descriptive bit in the auxiliary error register. The display's
ERR
annunciator illuminates whenever an error register bit is set. You can access
the error registers using ERRSTR? (both registers'), ERR? (error register only)
or AUXERR? (auxiliary error register only).
–
NOTE
: The internal self-test checks all calibration constants and verifies they
are within the lower and upper limits.
–
Related commands
: AUXERR?, ERR?, ERRSTR?
Example
OUTPUT 722;"TEST"
!RUNS SELF-TEST
Содержание 3458A
Страница 1: ...Keysight 3458A Multimeter Calibration Manual ...
Страница 10: ...10 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 14: ...14 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 16: ...16 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 18: ...18 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 39: ...Adjustment Procedures 3 Keysight 3458A Calibration Manual 39 Figure 3 1 4 terminal short ...
Страница 52: ...3 Adjustment Procedures 52 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 59: ...Performance Verification Tests 4 Keysight 3458A Calibration Manual 59 Figure 4 1 General test procedure ...
Страница 109: ...Command Summary 5 Keysight 3458A Calibration Manual 109 Related commands ACAL CAL SECURE ...
Страница 116: ...A Specifications 116 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...