5
Command Summary
90
Keysight 3458A Calibration Manual
CAL?
Description
Calibration query
. Returns a string containing one of four values for the
calibration constant specified; the initial (nominal) value, low limit, high limit, or
actual value of the specified constant. The returned string also contains a
description of the constant. This command is in the full command menu; it is not
in the short command menu.
Syntax
CAL? const_id [,cal_item]
cal_item
Specifies which of the four calibration constant values is to be returned. The
cal_item
parameter choices are:
cal_item
Description
0
Initial (nominal) value
1
Actual value
[a]
[a] The default for
cal_item
is the actual value.
3
Upper limit
5
Lower limit
Содержание 3458A
Страница 1: ...Keysight 3458A Multimeter Calibration Manual ...
Страница 10: ...10 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 14: ...14 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 16: ...16 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 18: ...18 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 39: ...Adjustment Procedures 3 Keysight 3458A Calibration Manual 39 Figure 3 1 4 terminal short ...
Страница 52: ...3 Adjustment Procedures 52 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 59: ...Performance Verification Tests 4 Keysight 3458A Calibration Manual 59 Figure 4 1 General test procedure ...
Страница 109: ...Command Summary 5 Keysight 3458A Calibration Manual 109 Related commands ACAL CAL SECURE ...
Страница 116: ...A Specifications 116 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...