3
Adjustment Procedures
50
Keysight 3458A Calibration Manual
11
Set the synthesizer to output 3 V rms at 100 kHz. Execute the
SCAL 1E5 command. The multimeter automatically performs the adjustment.
When the adjustment is complete, the multimeter returns to displaying DC
voltage readings.
12
Without changing the synthesizer settings, execute the SCAL 10 command as
shown on Line B of the Adjustment Record.
13
Set the synthesizer to the voltage and frequency shown on Line C of the
Adjustment Record. Execute the SCAL command as shown on Line C of the
Adjustment Record.
14
Repeat step 13 for each synthesizer setting and SCAL command shown on
Lines D through H on the Adjustment Record.
15
Disconnect all equipment from the multimeter.
16
Execute the ACAL AC command.
Содержание 3458A
Страница 1: ...Keysight 3458A Multimeter Calibration Manual ...
Страница 10: ...10 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 14: ...14 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 16: ...16 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 18: ...18 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 39: ...Adjustment Procedures 3 Keysight 3458A Calibration Manual 39 Figure 3 1 4 terminal short ...
Страница 52: ...3 Adjustment Procedures 52 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 59: ...Performance Verification Tests 4 Keysight 3458A Calibration Manual 59 Figure 4 1 General test procedure ...
Страница 109: ...Command Summary 5 Keysight 3458A Calibration Manual 109 Related commands ACAL CAL SECURE ...
Страница 116: ...A Specifications 116 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...