Keysight N774xC Series Multiport Power Meter User’s Guide
3
Definition of Terms
Port separation (static)
Specifies the difference between the power P applied to one port and the
measured power D on any other port with no applied power (dark).
Excluding "Noise pp (dark)" (see,
on page 44) and "Drift
(dark)" (see,
Powers expressed in dBm, port separation S expressed in dB,
S = P – D
Figure 12
Port separation (static)
Conditions:
Applied power level as specified.
Port separation, dynamic
Specifies the difference between step power input P
step
(with positive or
negative slope) applied to one port and the maximum measured power
change D
chng
(positive or negative) on any other port without applied
power (dark). Excluding "Noise pp (dark)" (see,
Powers expressed in dBm, dynamic port separation SD expressed in dB.
SD = P
step
- D
chng
Conditions:
Power step height as specified (1 mW).
Power range
Specifies the range from the smallest input power that causes a significant
change of the measured (displayed) power to the highest power for which
the specifications apply if not differently stated.