Probe Calibration
2
N2797A User’s Guide
29
Figure 11
E2655B PV Fixture with 50 Ohm Termination
• Position the probe so that the signal lead touches the PV fixture’s center conductor next
to the dialectric area. The ground lead should touch the PV fixture’s ground plane. On
90000 X-series oscilloscopes, use an N5442A adapter when instructed to connect the
probe to one of the oscilloscope’s channel inputs.
CAUTION
To avoid damaging the oscilloscope’s
Cal Out
connector, do not apply force to the PV fixture. Light probe
contact is all that is needed for the calibration.
CAUTION
NEVER solder a probe tip to the thick-film gold. The gold will immediately dissolve into the solder and
disappear.
NOTE
If you are using the browser tip, it is recommended that you use the N2787A 3D Probe
Positioner to hold the probe in place.
NOTE
You can check that the probe leads are correctly connected by pressing the front panel
autoscale button. A stable step should be shown on the screen. Pressing autoscale will close
the Probe Calibration dialog box.