44
N2797A User’s Guide
5
Performance Verification
Figure 19
Forward and Reverse Setup
18
Press the
transmission
screen key.
19
Press the
do both fwd and reverse
screen key.
20
Wait until the VNA beeps four times indicating that it has completed the task.
21
Press the
isolation
screen key.
22
Press the
omit isolation
screen key.
23
Press
done 2 port cal
screen key.
24
Set the VNA's averaging to off.
25
Save the reference plane cal by pressing the
[save recall]
key then the
[save state]
key.
26
You may change name if you wish.
27
Press the
[scale reference]
key. Then set the scale to 1 dB per division and the reference
position for 7 divisions.
28
Set reference value for 0 dB.
29
Press the
[measure]
key.
30
Press the
s21
screen key.
31
Ensure s21 response on screen is flat out to 1.5 GHz.