Performance Verification
5
N2797A User’s Guide
41
Figure 16
Probe Connected to Probe Adapter (
not to scale
)
Calibrating a Reference Plane
To get a reliable measurement from the VNA you must calibrate a reference plane so that
the VNA knows where the probe under test is located along the transmission line.
7
On the VNA, press the
[Cal]
key.
8
Press the
cal menu
screen key.
9
Press the
full 2 port
screen key.
10
Connect the PV fixture to the
Port 1
on the VNA using a high quality SMA cables. Connect
the cable to the pincher side of PV fixture as shown in
11
The calibration reference plane is at the other end of PV fixture.