3
Testing 16860 Performance
36
Keysight 16860 Series Portable Logic Analyzer Service Guide
Create Test Configuration Files
The test consists of multiple setups to test all the various analyzer modes. To make this easier, 4
configuration files will be created. The configuration files will then be used as starting points for the
measurement setups.
The analyzer has two state modes: Multiple Clocks and Single Clock. Two configuration files will be
created for each of these modes. One for finding the maximum clock frequency (no Marker) and
another to verify that the data is correct (with Marker).
Create Configuration File 1 - Multiple Clocks (no Marker)
1 Ensure that there is a U4203A connected to Pods 1 and 2. If Pods 3 and 4 exist, connect a
U4203A to these pods also.
2 In the Keysight Logic and Protocol Analyzer application, choose
File
>
New
. This puts the logic
analyzer into its initial state.
3 From the main drop down menu, select
Setup
then select
Bus/Signal
. The
Analyzer Setup
dialog is
displayed.
4 In the Analyzer Setup dialog, click the
Threshold
button for Pod 1.
5 The
Threshold Setting
dialog is displayed. Set the Threshold to
0 V
.
Содержание 16860 Series
Страница 1: ...Keysight 16860 Series Portable Logic Analyzers Service Guide...
Страница 8: ...8 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Страница 12: ...12 Keysight 16860 Series Portable Logic Analyzer Service Guide Contents...
Страница 32: ...3 Testing 16860 Performance 32 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Страница 34: ...3 Testing 16860 Performance 34 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Страница 104: ...5 Troubleshooting 104 Keysight 16860 Series Portable Logic Analyzers Service Guide...
Страница 136: ...7 Replaceable Parts 136 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Страница 138: ...Index 138 Keysight 16860 Series Portable Logic Analyzer Service Guide...