3
Testing 16860 Performance
42
Keysight 16860 Series Portable Logic Analyzer Service Guide
Create Configuration File 3 - Single Clock (no Marker)
1 Start with the
MultipleClocks_NoMarker
configuration file. Select
File
then
Open
. In the
Open
dialog
select the file and then open. If asked about saving the current configuration, click
No
.
2 Ensure that there is a U4203A connected to Pods 1 and 2. If Pods 3 and 4 exist, connect a
U4203A to these pods also. The Pod 1 leads should be connected to the generator as described
above.
3 Click the
Analyzer Setup > Sampling
tab.
4 In the
State Options
section, change the
Sampling
options to
Single Clock
.
5 In the clock selection area, change the Pod 1 clock to
Both Edges
.
6 Close the dialog by clicking
OK
.
7 Save the configuration by selecting
File
then
Save As …
.
Содержание 16860 Series
Страница 1: ...Keysight 16860 Series Portable Logic Analyzers Service Guide...
Страница 8: ...8 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Страница 12: ...12 Keysight 16860 Series Portable Logic Analyzer Service Guide Contents...
Страница 32: ...3 Testing 16860 Performance 32 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Страница 34: ...3 Testing 16860 Performance 34 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Страница 104: ...5 Troubleshooting 104 Keysight 16860 Series Portable Logic Analyzers Service Guide...
Страница 136: ...7 Replaceable Parts 136 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Страница 138: ...Index 138 Keysight 16860 Series Portable Logic Analyzer Service Guide...