3
Testing 16860 Performance
72
Keysight 16860 Series Portable Logic Analyzer Service Guide
15 Click the
Stop
toolbar button to stop the data acquisition.
16 Note the generator frequency setting. This will be used in the next section to verify the data rate.
Single Clock Both Edges - Pod 1 Data - Setup for Maximum Data Rate
1 On the 81134A, select the channel 2 setup screen. This is the data channel.
2 Set the Channel 2 Freq. Divider to 1.
3 Load the
SingleClock_WithMarker.ala
configuration file. If asked to save the current configuration
file, click
No
.
4 Verify that the Generator is set to the frequency found in the last section as a starting point.
5 Open the Sampling tab in the Analyzer Setup dialog by clicking the
Sampling Setup
Icon.
6 Adjust the sample position using the procedure described earlier by clicking the
Eye Scan: Sample
Position and Thresholds
button. Select the eye closest to -1 ns.
7 Close the dialog windows by clicking OK.
Single Clock Both Edges - Pod 1 Data - Measuring Maximum Data Rate
1 Select the
Run Repetitive
icon
. Let the logic analyzer run for about 1 minute. The analyzer will
acquire data and the Listing Window will continuously update. A marker search window will
appear and show progress.
2 If incorrect data is found, the following window will appear. The data in the listing window should
be A's & 5's.
Содержание 16860 Series
Страница 1: ...Keysight 16860 Series Portable Logic Analyzers Service Guide...
Страница 8: ...8 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Страница 12: ...12 Keysight 16860 Series Portable Logic Analyzer Service Guide Contents...
Страница 32: ...3 Testing 16860 Performance 32 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Страница 34: ...3 Testing 16860 Performance 34 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Страница 104: ...5 Troubleshooting 104 Keysight 16860 Series Portable Logic Analyzers Service Guide...
Страница 136: ...7 Replaceable Parts 136 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Страница 138: ...Index 138 Keysight 16860 Series Portable Logic Analyzer Service Guide...