Keysight 16860 Series Portable Logic Analyzers Service Guide
101
Troubleshooting
5
8 Adjust sampling positions:
a
Click the
Eye Scan: Thresholds and Sample Positions
button. The
Eyescan - Sample Positions and
Threshold Settings
dialog will appear.
b
In the “Buses/Signals” section of the Eyescan - Sample Positions and Threshold Settings
dialog, make sure the check box next to “My Bus 1” is checked.
c
Click
Edit,
select the
Do time scan only
option as the scan dimension and then select the
Lock
Thresholds
checkbox.
d
Select the
Run This Measurement
button in the
Eyescan - Sample Positions and Threshold Settings
dialog.
y