53
Testing 16860 Performance
3
17 Click the
Stop
toolbar button to stop the data acquisition.
18 Note the generator frequency setting this will be used in the next section to verify the data rate.
Pod 2 Clock - Setup for Maximum Data Rate
1 Verify that the Generator is set to the frequency found in the last section as a starting point.
2 Load the
MultipleClocks_WithMarker.ala
configuration file. If asked to save the current configuration
file, click
No
.
3 Open the
Bus/Signal Setup
dialog by clicking the
Setup
Icon in the tool bar.
4 Unassign the data bits from Pod1.
5 Assign bits 2, 6, 10, and 14 of Pod 2.
6 Open the
Sampling
tab.
7 In the clock assignment area, set Pod 1 Clock to
Don't Care
and set Pod 2 Clock to
Both Edges
.
8 Adjust the sample position using the procedure described earlier by clicking the
Eye Scan: Sample
Position and Thresholds
button.
Содержание 16860 Series
Страница 1: ...Keysight 16860 Series Portable Logic Analyzers Service Guide...
Страница 8: ...8 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Страница 12: ...12 Keysight 16860 Series Portable Logic Analyzer Service Guide Contents...
Страница 32: ...3 Testing 16860 Performance 32 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Страница 34: ...3 Testing 16860 Performance 34 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Страница 104: ...5 Troubleshooting 104 Keysight 16860 Series Portable Logic Analyzers Service Guide...
Страница 136: ...7 Replaceable Parts 136 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Страница 138: ...Index 138 Keysight 16860 Series Portable Logic Analyzer Service Guide...