3
Testing 16860 Performance
26
Keysight 16860 Series Portable Logic Analyzer Service Guide
Equipment Required for the Performance Test
The following equipment is required for the performance test procedure.
Table 2 Equipment Required
Equipment
Critical Specification
Recommended Model/Part
Pulse Generator
(Qty: 1)
≥
800 MHz, two channels, differential outputs,
150-180 ps rise/fall time (if faster, use transition
time converters)
Keysight 81134A or equivalent
150 ps Transition Time Converter (Qty 4)
Required if pulse generator's rise time is less than
150 ps. (Pulse generator conditions: Voffset=1V,
δ
V=250 mV.) Required for 81134A
Keysight 15435A
Flying Lead Probe Set (Qty 2)
A combination of U4201A & E5382A can be used.
Keysight U4203A
SMA/Flying Lead test connectors, (f) SMA to (f) SMA
to Flying Lead Probe (Qty 4)
no substitute