BHK-MG (OPR) 032816
3-33
If digital programming is at zero, analog programming requires a voltage between zero and
–10V d-c
to be applied to terminal 9 (TB2 for voltage, TB3 for current). If digital program-
ming is combined with analog programming, the voltage range for the signal applied to ter-
minal 9 is:
Voltage programming: TB2-9: 0 to
Current programming: TB3-9: 0 to
where:
Eout
dp
= digitally programmed output voltage/voltage limit
Iout
dp
= digitally programmed output current/current limit
The uncommitted amplifiers can be configured to produce the required voltage from a number of
analog sources. Some basic analog configurations are presented in the following paragraphs:
•
External resistance
•
Low impedance voltage source
•
Grounded voltage source
•
High impedance voltage source
•
Current source
Different analog sources can be used simultaneously; e.g., resistance can be sued to control
voltage programming while a grounded voltage source controls current programming. Analog
programming can also be combined with digital programming (either locally via the keypad or
remotely via the GPIB bus) since the two programming methods are summed together. For
example, for the BHK 500-0.4MG Power Supply, if the output is programmed digitally to 200V,
and an analog signal which would produce a 100V output is applied to the uncommitted ampli-
fied, the output would go to 300V (assuming the value of the load and the programmed value of
output current allow the power supply to remain in voltage mode).
NOTE: It is important to remember that
BOTH
voltage and current channels must be pro-
grammed (e.g., if analog programming of voltage mode is used, the current limit
must
be programmed by either analog or digital programming).
The availability of the two uncommitted amplifiers, ±10V d-c reference voltages and the two
summing amplifiers, means that many configurations are possible for analog control of the
power supply.
3.6.1
ANALOG PROGRAMMING WARNINGS AND CAUTIONS
THE POWER SUPPLY PRODUCES DANGEROUS VOLTAGES WHICH CAN BE LETHAL.
ALWAYS OBSERVE THE FOLLOWING PRECAUTIONS.
a. Always connect the chassis of the power supply to a good AC ground (earth) con-
nected to OUTPUT TERMINALS terminal strip TB1, terminal 4.
NOTE: All connections must be tight, whether at the terminal strips of the power supply or at
external equipment.
10
E o u t
d p
–
10
E o m a x
----------------------
V
–
10
I
o u t
d p
–
10
I
o m a x
--------------------
V
–
WARNING
Содержание BHK 1000-0.2MG
Страница 2: ......
Страница 6: ......
Страница 17: ...BHK MG032816 xi FIGURE 1 1 BHK MG SERIES PROGRAMMABLE POWER SUPPLY ...
Страница 18: ...xii BHK MG032816 ...
Страница 24: ...1 6 BHK MG OPR 032816 FIGURE 1 2 BHK SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 1 OF 2 ...
Страница 25: ...BHK MG OPR 032816 1 7 FIGURE 1 2 BHK SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 2 OF 2 ...
Страница 46: ......
Страница 101: ...BHK MG OPR 032816 3 55 FIGURE 3 20 SERIES AUTOMATIC CONFIGURATION ...
Страница 109: ...BHK MG OPR 032816 3 63 FIGURE 3 23 PARALLEL AUTOMATIC CONFIGURATION ...
Страница 114: ...3 68 BHK MG OPR 032816 FIGURE 3 25 PARALLEL MASTER SLAVE CURRENT MODE CONFIGURATION ...
Страница 116: ......
Страница 154: ......