IEEE-488 Bus Overview
C-8
C.5
Interface function codes
The interface function codes, which are part of the IEEE-488
standards, define an instrument’s ability to support various
interface functions and should not be confused with pro-
gramming commands found elsewhere in this manual. The
interface function codes for the Model 708A are listed in
Table C-6.
The codes define Model 708A capabilities as follows:
• SH (Source Handshake Function) — SH1 defines the
ability of the instrument to initiate the transfer of mes-
sage/data over the data bus.
Table C-6
Model 708A interface function codes
Code
Interface function
SH1
AH1
T5
L4
SR1
RL1
PP0
DC1
DT1
C0
E1
TE0
LE0
Source Handshake capability
Acceptor Handshake capability
Talker (basic talker, talk-only, serial poll, unad-
dressed to talk on LAG)
Listener (basic listener, unaddressed to listen on
TAG)
Service Request capability
Remote/Local capability
No Parallel Poll capability
Device Clear capability
Device Trigger capability
No Controller capability
Open collector bus drivers
No Extended Talker capability
No Extended Listener capability
• AH (Acceptor Handshake Function) —
AH1 defines
the ability of the instrument to guarantee proper recep-
tion of message/data transmitted over the data bus.
• T (Talker Function) — The ability of the instrument to
send data over the bus to other devices is provided by
the T function. Instrument talker capabilities (T5) exist
only after the instrument has been addressed to talk.
• L (Listener Function) — The ability of the instrument
to receive device-dependent data over the bus from
other devices is provided by the L function. Listener
capabilities (L4) of the instrument exist only after it has
been addressed to listen.
• SR (Service Request Function)
—
SR1 defines the
ability of the instrument to request service from the
controller.
• RL (Remote/Local Function) — RL1 defines the ability
of the instrument to be placed in the remote or local
modes.
• PP (Parallel Poll Function) — The instrument does not
have parallel polling capabilities (PP0).
• DC (Device Clear Function) — DC1 defines the ability
of the instrument to be cleared (initialized).
• DT (Device Trigger Function) — DTI defines the abil-
ity of the Model 708A to have readings triggered.
• C (Controller Function) — The instrument does not
have controller capabilities (C0).
• E (Bus Driver Type) — The instrument has open-
collector bus drivers (E1).
• TE (Extended Talker Function) — The instrument does
not have extended talker capabilities (TE0).
• LE (Extended Listener Function) — The instrument
does not have extended listener capabilities (LE0).
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание 708A
Страница 135: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 136: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 137: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 138: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 139: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 145: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 146: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 147: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 148: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 149: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 150: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...