Calibration Command Reference
3-9
:OHM200
(CALibration:PROTected:DC:OHM200)
Purpose
To program the 200
Ω
comprehensive calibration step.
Format
:cal:prot:dc:ohm200 <Cal_resistance>
Parameter
<Cal_resistance> = 95 to 205 [
Ω
]
Description
:OHM200 programs the 200
Ω
comprehensive calibration step. The allowable range of the
calibration resistance parameteris from 95
Ω
to 205
Ω
. Use the 200
Ω
value whenever possible,
or the closest possible value (for example, 190
Ω
, which is the closet value available on many
calibrators).
Example
:CAL:PROT:DC:OHM200 190
Program 200
Ω
step.
:OHM20
(CALibration:PROTected:DC:OHM20)
Purpose
To program the 20
Ω
comprehensive calibration step.
Format
:cal:prot:dc:ohm20 <Cal_resistance>
Parameter
<Cal_resistance> = 9.5 to 20.5 [
Ω
]
Description
:OHM20 programs the 20
Ω
comprehensive calibration step. The allowable range of the
calibration resistance parameter is from 9.5
Ω
to 20.5k
Ω
. Use the 20
Ω
value whenever possible,
or the closest possible value (for example, 19
Ω
, which is the closet value available on many
calibrators).
Example
:CAL:PROT:DC:OHM20 19
Program 20
Ω
step.
:A200U
(CALibration:PROTected:DC:A200U)
Purpose
To program the 200µA comprehensive calibration step.
Format
:cal:prot:dc:a200u <Cal_current>
Parameter
<Cal_current> = 95E-6 to 205E-6 [A]
Description
:A200U programs the 200µA comprehensive calibration step. The allowable range of the cali-
bration current parameter is from 95µA to 205µA. Use the 200µA value whenever possible for
best results.
Example
:CAL:PROT:DC:A200U 200E-6
Program 200µA step.
Содержание 2002
Страница 1: ...Model 2002Multimeter Calibration Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Страница 9: ...iv...
Страница 11: ...vi...
Страница 55: ...Calibration 2 26...
Страница 70: ...Specifications A 1 A Specifications...
Страница 83: ...A 14 Specifications...
Страница 101: ...Calibration Programs B 18...
Страница 113: ......