![Keithley 2002 Скачать руководство пользователя страница 81](http://html1.mh-extra.com/html/keithley/2002/2002_calibration-manual_3975661081.webp)
A-12 Specifications
Operating Speed
Function Change Speed
1
Typical delay before measurement initiation after making a function change.
From
To
Function
Function
Range
Time
Any except 4W
Ω
, Temp
DCV
Any
4.6 ms
4W
Ω
, Temp
Any
7.6 ms
Any
ACV
Any
574
ms
ACV, DCV, 2W
Ω
, Freq
DCI
Any
7.1 ms
4W
Ω
, Temp
Any
10
ms
ACI
Any
22
ms
Any
ACI
Any
523
ms
Any except 4W
Ω
, Temp
2W
Ω
20
Ω
to 2k
Ω
4.7 ms
20k
Ω
15
ms
200k
Ω
27
ms
2M
Ω
103
ms
20M
Ω
153
ms
200M
Ω
, 1G
Ω
253
ms
4W
Ω
, Temp
2W
Ω
20
Ω
to 2k
Ω
7.7 ms
20k
Ω
18
ms
200k
Ω
30
ms
2M
Ω
105
ms
20M
Ω
157
ms
200M
Ω
, 1G
Ω
256
ms
Any
4W
Ω
20
Ω
to 2k
Ω
7.7 ms
20k
Ω
18
ms
200k
Ω
30
ms
2M
Ω
105
ms
Any except ACV, ACI
Freq
5
Any
60
ms
ACV, ACI
Any
573
ms
Any
Temp
Any
7.6 ms
Range Change Speed
1
Typical delay before measurement initiation after making a
range change.
Function
From
To
Time
DCV
Any
Any
5.2 ms
ACV
Any
Any
559
ms
DCI
Any
Any
7.6 ms
ACI
Any
Any
503
ms
2W
Ω
Any
20
Ω
to 2k
Ω
5.2 ms
Any
20k
Ω
15
ms
Any
200k
Ω
27
ms
Any
2M
Ω
103
ms
Any
20M
Ω
153
ms
Any
200M
Ω
, 1G
Ω
253
ms
4W
Ω
Any
20
Ω
to 2k
Ω
5.2 ms
Any
20k
Ω
15
ms
Any
200k
Ω
27
ms
Any
2M
Ω
103
ms
Trigger Speed (External Trigger or Trigger-Link)
Autozero Off
Autozero On
Trigger Latency:
< 2
µ
s
1.2 ms typical
Trigger Jitter:
±
0.5
µ
s
GPIB Data Formatting Transmission Time
2
Readings
Readings with
Only
Time Stamp
Format
Time
Rdg./s
Time Rdg./s
DREAL
(Double precision real)
0.51 ms
1961
3.1 ms
323
SREAL
(Single precision real)
0.38 ms
2632
3.3 ms
303
ASCII
6.2 ms
161
10.2 ms
98
Single Function Scan Speed
3
(Internal Scanner)
2W
Ω
4W
Ω
RTD Temp
DCV (20V)
(2k
Ω
)
(2k
Ω
)
ACV
Freq
TC Temp
(2-Wire)
Time
Rate
Time
Rate
Time
Rate
Time
Rate
Time
Rate
Time
Rate
Time
Rate
per
(Chan./
per
(Chan./
per
(Chan./
per
(Chan./
per
(Chan./
per
(Chan./
per
(Chan./
TYPE
Chan.
second)
Chan. second)
Chan. second)
Chan.
second)
Chan.
second)
Chan.
second)
Chan.
second)
Ratio or Delta
4
(2 channels)
8.2 ms
122
8.5 ms 118
18.8 ms
53
Fast Scan
(using solid state
channels)
8.2 ms
122
6.3 ms 159
501 ms
2
559 ms
1.8
12.8 ms
78
Normal Scan
14 ms
71
11.4 ms
88
14.4 ms
69
506 ms
2
564 ms
1.8
17.2 ms
58
43 ms
23
Operating Speed Notes
1 For display off, 0.01 power line cycles, autorange off,
digital filter off, autozero on, offset compensation off.
Display on may impact time by 3% worst case. To
eliminate this impact, press ENTER (hold) to freeze
display.
2 Using 386/33 computer, average time for 1000 readings,
byte order swapped, display off.
3 For on-scale readings, no trigger delays, display off, 0.01
power line cycles, autorange off, digital filter off, offset
compensation off, autozero off.
4 Ratio and delta functions output one value for each pair
of measurements.
5 Based on 100kHz input frequency.
Содержание 2002
Страница 1: ...Model 2002Multimeter Calibration Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Страница 9: ...iv...
Страница 11: ...vi...
Страница 55: ...Calibration 2 26...
Страница 70: ...Specifications A 1 A Specifications...
Страница 83: ...A 14 Specifications...
Страница 101: ...Calibration Programs B 18...
Страница 113: ......