Calibration Command Reference
3-7
:V2
(:CALibration:PROTected:DC:V2)
Purpose
To program the +2V comprehensive calibration step.
Format
:cal:prot:dc:v2 <Cal_voltage>
Parameter
<Cal_voltage> = 0.95 to 2.05 [V]
Description
:V2 programs the +2V DC comprehensive calibration step. The allowable range of the calibra-
tion voltage parameter is from 0.95 to 2.05V, but 2V is recommended for best results.
Example
:CAL:PROT:DC:V2 2
Program 2V step.
:V20
(:CALibration:PROTected:DC:V20)
Purpose
To program the +20V DC comprehensive calibration step.
Format
:cal:prot:dc:v20 <Cal_voltage>
Parameter
<Cal_voltage> = 9.5 to 20.5 [V]
Description
:V2 programs the +20V DC comprehensive calibration step. The allowable range of the calibra-
tion voltage parameter is from 9.5 to 20.5V, but 20V is recommended for best results.
Example
:CAL:PROT:DC:V20 20
Program 20V step.
:OHM1M
(CALibration:PROTected:DC:OHM1M)
Purpose
To program the 1M
Ω
comprehensive calibration step.
Format
:cal:prot:dc:ohm1m <Cal_resistance>
Parameter
<Cal_resistance> = 475E3 to 1.025E6 [
Ω
]
Description
:OHM1M programs the 1M
Ω
comprehensive calibration step. The allowable range of the cali-
bration resistance parameter is from 475k
Ω
to 1.025M
Ω
. Use the 1M
Ω
value whenever possi-
ble, or the closest possible value.
Example
:CAL:PROT:DC:OHM1M 1E6
Program 1M
Ω
cal step.
Содержание 2002
Страница 1: ...Model 2002Multimeter Calibration Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Страница 9: ...iv...
Страница 11: ...vi...
Страница 55: ...Calibration 2 26...
Страница 70: ...Specifications A 1 A Specifications...
Страница 83: ...A 14 Specifications...
Страница 101: ...Calibration Programs B 18...
Страница 113: ......